Posted on

Comprehensive User Guide for Hach DR1010 COD Determinator

Preface: The Importance of COD Determination Technology and an Overview of the Instrument

Chemical Oxygen Demand (COD) is a crucial indicator in water quality monitoring, reflecting the extent of water pollution caused by reducing substances. The Hach DR1010 COD Determinator, a professional water quality analysis instrument, is widely used in environmental monitoring, sewage treatment, and industrial wastewater testing. This guide aims to comprehensively analyze the operational procedures, functional features, maintenance, and troubleshooting methods of the DR1010 based on the user manual, helping users obtain accurate and reliable test results.

Developed by Hach Company, the DR1010 COD Determinator is controlled by a microprocessor and features an LED light source, suitable for laboratory or on-site measurements. It has four built-in COD test programs, supports user-created curves, and can store up to 40 user programs. The instrument offers flexible power supply options, including a 6V adapter or four AA alkaline dry batteries, operates within a temperature range of 0 to 50°C, and meets the IP41 protection standard.

Chapter 1: Instrument Structure and Function Details

1.1 Instrument Composition and Standard Accessories

The standard configuration of the DR1010 COD Determinator includes:

  • Power adapter (Product No.: 9185600)
  • Data transfer cable (RS232 port, black)
  • Document bag (containing operation manual, method manual, and certificate of conformity)

Optional accessories:

  • COD test tubes (16mm × 100mm, with tube caps)
  • Data printing cable (RS232 port, gray)
  • DRB200 digestor
  • Bottle-top dispensers
  • Pipettes

1.2 Instrument Technical Parameters

  • Wavelength range: 420nm and 610nm dual wavelengths
  • Wavelength accuracy: ±1nm
  • Photometric measurement linearity: ±0.002A (0-1A)
  • Photometric measurement repeatability: ±0.005A (0-1A)
  • Light source: LED
  • Detector: Silicon photodiode
  • Data display: Four-digit LCD, 1.5 cm character height
  • Readout modes: % transmittance, absorbance, concentration
  • External output: RS232 serial port
  • Power supply: 190~240VAC/50Hz adapter or four AA alkaline batteries
  • Instrument dimensions: 24.0 × 19.8 × 12.0 cm
  • Instrument weight: 2 kg
  • Operating temperature: 0 to 50°C
  • Storage temperature: -20 to 60°C

1.3 Keyboard Function Details

Program Selection Keys:

  • High-range 2h: Selects the high-range two-hour digestion method; acts as the number key 7 in numeric mode.
  • Low-range 2h: Selects the low-range two-hour digestion method; acts as the number key 4 in numeric mode.
  • High-range rapid: Selects the high-range 15-minute digestion method; acts as the number key 1 in numeric mode.
  • Low-range rapid: Selects the low-range 15-minute digestion method; acts as the number key 1 in numeric mode.

Function Keys:

  • Print: Prints current data; acts as the number key 8 in numeric mode.
  • Save: Stores the current reading; acts as the number key 5 in numeric mode.
  • Historical data: Retrieves stored sample data; acts as the number key 2 in numeric mode.
  • Zero: Uses the current sample blank for zero adjustment; acts as the number key 0 in numeric mode.
  • Setup: Enters the setup menu; acts as the number key 9 in numeric mode.
  • Time/Date: Displays the current time or date; acts as the number key 6 in numeric mode.
  • Unit conversion: Converts between concentration, absorbance, and % transmittance; acts as the number key 3 in numeric mode.
  • Read: Reads and displays the sample concentration; inputs decimal points or switches between positive and negative signs in numeric mode.
  • Return: Cancels the current input or selection.
  • △/▽: Scrolls up and down within the menu.
  • Enter: Selects a menu item or accepts an input value.

Chapter 2: Initial Instrument Setup and Calibration

2.1 Battery Installation and Power Management

  • Turn the instrument over and ensure the sample cell is empty.
  • Open the battery compartment cover and install four AA alkaline batteries according to the markings.
  • Re-cover the battery compartment and turn the instrument back to its upright position.

Important Tips:

  • Use alkaline batteries. Do not use rechargeable Ni-Cd batteries.
  • Replace all batteries when changing them.
  • When the battery level is low, the LOW BATTERY icon will be displayed. Replace the batteries promptly.
  • It is recommended to remove the batteries if the instrument is not used for an extended period.

2.2 Date and Time Setup

Date Setup:

  • Press the “Setup” key to enter the SETUP menu.
  • Select the DATE option and input the four-digit year, month, and day.
  • Press the “Enter” key to confirm.

Time Setup:

  • In the SETUP menu, select the TIME option.
  • Input the time in 24-hour format.
  • Press the “Enter” key to confirm.

2.3 Proper Use of Sample Tubes

  • Wipe the outer surface of the sample tube with a lint-free cloth.
  • Insert the tube into the instrument’s tube holder, with the HACH logo facing the display.
  • Ensure consistent insertion direction for each measurement.
  • Check that the sample tube is clean and free of scratches before measurement.

Chapter 3: Detailed Instrument Operation Procedures

3.1 Basic Measurement Steps

Determinator Setup:

  • Upon startup, the instrument automatically enters the program used last time.
  • Press the corresponding program key to select a program and press the “Enter” key to confirm.

Sample Preparation:

  • Prepare the zero solution and the sample to be tested according to the program instructions.

Instrument Zeroing:

  • Place the blank solution in the sample cell.
  • Close the cover and press the “Zero” key.
  • When the instrument displays 0 and the READ icon appears, measurement can begin.

Sample Measurement:

  • Place the sample to be tested in the holder.
  • Close the cover and press the “Read” key.
  • The display shows the measurement result.
  • Press the “Unit conversion” key to switch the display mode.

3.2 Standard Curve Adjustment Method

  • Prepare standard solutions.
  • Measure the standard solutions as samples in the program.
  • After obtaining the readings, press the “Setup” key and scroll to the “STD” setting item.
  • Input the actual concentration of the standard solution and press the “Enter” key.

Notes:

  • Consider sample interference before adjustment.
  • After adjustment, test multiple concentration standard solutions to verify the applicability of the curve.
  • If the input calibration value is out of range, the instrument will emit a beep to indicate an error.

3.3 Data Storage and Retrieval

Data Storage:

  • After the measurement result is displayed, press the “Save” key.
  • The display shows the next available storage sequence number.
  • Press the “Enter” key to accept or input a specific sequence number.

Data Retrieval:

  • Press the “Historical data” key to enter the RECALL menu.
  • Use the “▽” or “△” key or numeric keys to select the sample sequence number.
  • Press the “Enter” key to display the stored data.

Chapter 4: Advanced Function Applications

4.1 User Program Creation Method

  • Press the “Setup” key and select the USER option.
  • Input the program number to be created (20-59).
  • Select the wavelength.
  • Prepare standard solutions and perform zero adjustment on the instrument.
  • Measure the absorbance values of the standard solutions.
  • Repeat the steps to complete the input of all standard points.
  • Press the “Return” key and select to store the program.

Key Points:

  • A minimum of 2 data points and a maximum of 12 are required.
  • At 420nm, the absorbance should decrease as the concentration increases.
  • At 610nm, the absorbance should increase as the concentration increases.
  • The instrument will ignore identical absorbance values and emit a beep.

4.2 Data Transmission and Printing

Printer Connection:

  • Connect the instrument and the printer using the gray data printing cable.
  • Press the “Print” key to manually initiate printing.

Computer Connection:

  • Connect the instrument and the computer using the black data transfer cable.
  • Set the super terminal parameters.
  • Start the text capture function.
  • Press the “Print” key to transmit data to a text file.

4.3 Batch Data Processing

  • Print all data: Select PRINT ALL in the SETUP menu.
  • Delete all data: Select ERASE ALL in the SETUP menu.
  • Data export: Transfer all data to a computer through the RS232 interface.

Chapter 5: Instrument Maintenance and Troubleshooting

5.1 Daily Maintenance Points

Cleaning and Maintenance:

  • Wipe the instrument’s outer shell with a damp cloth.
  • Promptly clean up any spilled reagents.
  • Clean the sample cell holder with a cotton swab.
  • Wipe the outer surface of the sample cell with lens paper or a soft, lint-free cloth.

Battery Management:

  • Replace low-battery cells promptly.
  • Remove the batteries if the instrument is not used for an extended period.
  • Reset the date and time after replacing the batteries.

Storage Conditions:

  • Storage temperature: -20 to 60°C
  • Relative humidity: Below 80% (at 40°C)
  • Avoid strong electromagnetic field environments.

5.2 Common Fault Exclusion

Error Codes and Solutions:

  1. Unable to set the instrument. Contact Hach customer service.
  2. Unable to read program data. Contact Hach customer service.
  3. Unable to write program data. Contact Hach customer service.
  4. Measurement battery error. Replace the batteries.
  5. Measurement A/D error. Contact Hach customer service.
  6. Measurement offset error. Check the installation of the light blocker.
  7. Low photometric intensity error. Check for light channel blockage or dilute the sample.
  8. Measurement value out of range. Confirm the installation of the instrument cover or contact customer service.

Other Common Problems:

  • Concentration out of range: Dilute the sample and re-measure.
  • Beep/error icon: Check the operational steps.
  • Low battery level: The LOW BATTERY icon is displayed. Replace the batteries promptly.

Chapter 6: Safety Regulations and Quality Assurance

6.1 Safety Operation Regulations

Hazard Levels:

  • Danger (DANGER): Situations that may lead to death or serious injury.
  • Caution (CAUTION): Situations that may lead to minor or moderate injury.
  • Note (NOTE): Information that requires special emphasis.

Key Safety Tips:

  • Review the Material Safety Data Sheet (MSDS) and be familiar with safety procedures when handling chemical samples.
  • The instrument should not be used for samples that are flammable or contain hydrocarbons.
  • Do not use Ni-Cd rechargeable batteries.
  • Do not open the instrument’s chassis without authorization.

6.2 Quality Assurance and Service Support

Quality Assurance:

  • Most products are guaranteed for at least one year from the shipping date.
  • The warranty covers defects in materials and manufacturing.

Repair Services:

  • Users should not attempt to repair any parts other than the batteries by themselves.
  • Contact an authorized Hach Company service center for repairs.

Chapter 7: Practical Application Tips and Experience Sharing

7.1 Best Practices for COD Measurement

Sample Handling Tips:

  • Ensure the sample is representative and mix it thoroughly before sampling.
  • Follow the digestion time and temperature requirements strictly.
  • Use reagents from the same batch for comparative measurements.

Methods to Reduce Errors:

  • Regularly verify the instrument’s accuracy using standard solutions.
  • Keep the sample tube clean.
  • Perform zero adjustment before each measurement.
  • Take the average of multiple measurements of the same sample.

7.2 Handling Special Application Scenarios

High-Salinity Sample Measurement:

  • May cause interference. It is recommended to conduct a spike recovery test.
  • Establish a specific calibration curve if necessary.

Low-Concentration Sample Measurement:

  • Use the low-range program to improve sensitivity.
  • Extend the measurement time or increase the sample volume.

Chapter 8: Instrument Verification and Compliance

8.1 Performance Verification Methods

Blank Test:

  • Measurement of ultrapure water should show 0mg/L COD.

Standard Sample Test:

  • Use COD standard solutions with known concentrations for verification.

Repeatability Test:

  • Measure the same sample multiple times and calculate the relative standard deviation.

Comparison Test:

  • Compare the results with standard methods or other instruments.

8.2 Compliance Certification

LED Safety:

  • Complies with EN60825-1 standard, Class 1 LED product.

Anti-Interference Characteristics:

  • Complies with EN 50082-1 general anti-interference standard.

EMC Electromagnetic Compatibility:

  • EN 61000-4-2 resistance to electrostatic discharge interference.
  • EN 61000-4-3 resistance to radiated RF electromagnetic field interference.
  • ENV 50204 resistance to digital telephone radiation.

Radio Frequency Emissions:

  • Complies with EN 55011 (CISPR 11) Class B emission limits.

Conclusion

The Hach DR1010 COD Determinator is a powerful and easy-to-use professional water quality analysis instrument. Through systematic learning of this guide, users should be able to master all the functions of the instrument, from basic operations to advanced applications. Correct operational methods and regular maintenance not only ensure the accuracy of measurement data but also extend the instrument’s service life. When encountering problems that cannot be resolved, users should promptly contact Hach Company’s professional technical service personnel to avoid improper operations that may cause instrument damage or data loss.

With the continuous improvement of environmental protection requirements, the importance of COD monitoring is becoming increasingly prominent. It is hoped that this guide will help users fully leverage the performance advantages of the DR1010 COD Determinator and provide reliable technical support for water quality monitoring and environmental protection work.

Posted on

User Manual and Operation Guide for Thermo Fisher FlashSmart Intelligent Elemental Analyzer (FlashSmart EA)

I. Instrument Overview and Basic Operations

1.1 Instrument Introduction

The Thermo Fisher FlashSmart Elemental Analyzer is a fully automated organic elemental analysis system that employs the dynamic combustion method (modified Dumas method) to determine nitrogen, carbon, hydrogen, and sulfur content. It measures oxygen content through high-temperature pyrolysis. This instrument can be configured with a single-channel or dual independent-channel system, and the MultiValve Control (MVC) module enables automatic dual-channel switching for analysis.

Main Technical Parameters:

  • Detector Type: Thermal Conductivity Detector (TCD)
  • Power Supply: 230V ± 10%, 50/60Hz, 1400VA
  • Dimensions: 50cm (height) × 59cm (width) × 58cm (depth)
  • Weight: 65kg
  • Maximum Operating Temperature: 1100℃
  • Gas Requirements: High-purity helium (carrier gas), oxygen (combustion aid), argon (for specific configurations)

1.2 Safety Precautions

Hazardous Operation Warnings:

  • High Voltage Risk: The instrument contains high-voltage components. Non-professionals are prohibited from opening the electrical compartment.
  • High-Temperature Surfaces: The furnace can reach temperatures up to 1100℃. Avoid contact during operation.
  • Gas Safety: Hydrogen use requires extreme caution, as concentrations as low as 4% pose an explosion risk.
  • Chemical Hazards: Wear protective gear when handling reaction tube packing materials and sample ashes.

Personal Protective Equipment (PPE) Requirements:

  • Eye Protection: Splash-resistant goggles
  • Hand Protection: White nitrile gloves (for chemicals)/heat-resistant gloves (for high-temperature operations)
  • Respiratory Protection: Dust masks
  • Body Protection: Lab coats + plastic aprons

1.3 Startup Preparation Procedure

Gas Connection:

  • Helium Inlet Pressure: 2.5bar (36psig)
  • Oxygen Inlet Pressure: 2.5-3bar (36-44psig)
  • Argon Inlet Pressure: 2.5bar (N/Protein configuration) or 4-4.5bar (NC Soils configuration)
  • Leak Testing: Perform on all gas lines.

Power Connection:

  • Confirm voltage stability at 230V ± 10%.
  • Ensure proper grounding; avoid sharing circuits with large motor equipment.

Software Installation:

  • System Requirements: Windows 7/8/10, at least 1GB hard drive space.
  • Install EagerSmart data processing software and drivers.

II. Calibration and Adjustment Procedures

2.1 Initial Setup

Hardware Configuration Steps:

  • Select Reaction Tube Configuration Based on Analysis Needs:
    • CHN Mode: Quartz reaction tube + chromium oxide/reduced copper/cobalt oxide packing.
    • CHNS Mode: Quartz reaction tube + copper oxide/electrolytic copper packing.
    • O Mode: Quartz reaction tube + nickel-plated carbon/quartz shavings packing.
    • N Mode: Dual reaction tubes in series + Plexiglas adsorption filter.
  • Install Autosampler:
    • MAS Plus Solid Autosampler: Up to 125-position sample tray.
    • AI 1310/AS 1310 Liquid Autosamplers: 8-position or 105-position sample trays.
  • Connect MVC Module (Dual-Channel Configuration):
    • Remove bypass panel from the rear.
    • Connect gas lines for left and right channels.
    • Configure dual MAS Plus autosamplers.

2.2 System Calibration

Three-Step Calibration Method:

  • Leak Testing:
    • Initiate automatic leak detection via software.
    • Acceptable Leak Rate: <0.1mL/min.
    • Use soapy water to locate leaks if detected.
  • Signal Baseline Adjustment:
    • Set TCD detector temperature constant (typically 40-120℃).
    • Adjust bridge voltage to 5V.
    • Baseline Drift: Should be <0.1mV/10min.
  • Standard Curve Establishment:
    • Use high-purity standards like acetanilide (nitrogen 16.09%, carbon 71.09%, hydrogen 6.70%).
    • Minimum Concentration Gradients: 5 points (recommended range: 0.1-5mg).
    • Correlation Coefficient (R²): Should be >0.999.

Calibration Frequency Recommendations:

  • Daily Use: Calibrate after each startup.
  • Continuous Analysis: Verify calibration every 50 samples.
  • After Consumable Replacement: Recalibration is mandatory.

2.3 Method Optimization

Parameter Adjustment Guidelines:

  • Oxygen Injection Time:
    • Regular Samples: 4-6 seconds.
    • Refractory Samples: Extend to 8 seconds.
    • High-Sulfur Samples: Add vanadium pentoxide as a combustion aid.
  • Furnace Temperature Settings:
    • Combustion Furnace: 950-1100℃.
    • Reduction Furnace: 840℃.
    • Pyrolysis Furnace (O Mode): 1060℃.
  • Carrier Gas Flow Rate:
    • Helium: 100-140mL/min.
    • Reference Gas: 30-50mL/min.

III. Routine Maintenance

3.1 Regular Maintenance Schedule

Maintenance Schedule Table:

Maintenance ItemFrequencyKey Operation Points
Reaction Tube RegenerationEvery 200 analysesEmpty packing material, incinerate at 550℃ for 2 hours.
Adsorbent ReplacementMonthlyActivate molecular sieve at 300℃, replace desiccant (silica gel) promptly.
Autosampler CleaningWeeklyUltrasonically clean tin/silver cups, inspect piston seals.
Chromatographic Column AgingQuarterlyAge at 280℃ with carrier gas for 8 hours.
Comprehensive System VerificationAnnuallyConducted by a professional engineer.

3.2 Key Component Maintenance

Reaction Tube Packing Guidelines:

  • Quartz Reaction Tubes:
    • Begin packing from the conical end.
    • Compact each layer with a dedicated tamping rod.
    • Separate layers with quartz wool.
    • Maintain total packing height at 80% of tube length.
  • HPAR Alloy Steel Reaction Tubes:
    • Must be used with crucibles.
    • Ensure uniform distribution of oxidation catalysts.
    • Use dedicated tools for installation/removal.

Adsorption Filter Maintenance:

  • Large (Plexiglas) Filters:
    • Packing sequence: Quartz wool → soda lime → molecular sieve → silica gel.
    • Pre-moisten soda lime with 0.5mL water.
  • Small (Pyrex) Filters:
    • Used in CHNS/O modes.
    • Packing: Quartz wool → anhydrous magnesium perchlorate.

3.3 Consumable Replacement Intervals

Recommended Replacement Intervals:

  • Quartz Wool: Replace when changing reaction tube packing.
  • Reduced Copper: Every 500 analyses.
  • Oxidation Catalyst: Every 300 analyses.
  • Nickel-Plated Carbon (O Mode): Every 150 analyses.
  • TCD Filament: Replace when baseline noise occurs.
  • Sealing O-Rings: Replace if leaks are detected or every 6 months.

IV. Troubleshooting and Solutions

4.1 Common Error Codes

Error Code Table:

CodeMeaningSolution
E01Left Furnace Temperature ExceededCheck thermocouple connection, restart system.
E04TCD Signal OverflowAdjust gain, verify carrier gas purity.
E12Safety Cutoff TriggeredCheck cooling fan, allow system to cool.
E25EFC-t Module Flow AbnormalityCheck for gas line blockages, clean filter.
E33Autosampler Communication FailureReconnect cables, verify port settings.

4.2 Typical Problem Resolution

Analysis Result Anomaly Investigation:

  • Low Nitrogen Results:
    • Check if reduced copper is失效 (discolored black).
    • Verify adequate oxygen injection.
    • Confirm complete sample combustion (observe flame).
  • Sulfur Peak Tailings:
    • Replace copper oxide packing layer.
    • Add vanadium pentoxide combustion aid.
    • Check chromatographic column connections for leaks.
  • Unstable Oxygen Results:
    • Verify nickel-plated carbon packing height (should be 60mm).
    • Confirm silver cup seal integrity.
    • Validate pyrolysis furnace temperature stability (±2℃).

Hardware Fault Handling:

  • Furnace Temperature Failure to Rise:
    • Check SSR solid-state relay status.
    • Measure transformer output voltage (should be 48V AC).
    • Confirm fuse integrity (AC 1112 board F1/F2).
  • Abnormal Gas Flow:
    • Clean EFC-t module filter.
    • Verify solenoid valve EV1-EV4 operation.
    • Calibrate flow sensors S1/S2.
  • TCD Baseline Drift:
    • Extend equilibration time to 2 hours.
    • Verify reference gas flow stability.
    • Replace aged filament.

4.3 Emergency Response Procedures

Safety Emergency Plan:

  • Gas Leak:
    • Immediately close cylinder main valve.
    • Activate laboratory ventilation system.
    • Avoid operating electrical equipment.
  • Furnace Overheating:
    • Trigger front panel emergency stop button.
    • Cut off main power supply.
    • Purge system with inert gas.
  • Abnormal Combustion:
    • Maintain system enclosure.
    • Direct exhaust through fume hood.
    • Do not cool directly with water.

V. Advanced Application Techniques

5.1 Special Sample Handling

Solutions for Challenging Samples:

  • High Inorganic Salt Samples:
    • Use quartz crucibles to prevent corrosion.
    • Reduce quartz wool between packing layers.
    • Increase oxygen injection pressure by 10%.
  • Volatile Liquids:
    • Utilize AI 1310 liquid autosampler.
    • Adsorb sample onto diatomaceous earth.
    • Preheat injection needle to 40℃.
  • Viscous Samples:
    • Grind with quartz sand for homogenization.
    • Use specially shaped tin cups.
    • Extend combustion time by 20%.

5.2 Data Quality Enhancement

Best Practice Recommendations:

  • Sample Preparation:
    • Homogenize to below 80 mesh.
    • Pre-dry samples with >5% moisture content.
    • Avoid fluorine-containing containers.
  • Weighing Techniques:
    • Use blank tin cups for calibration with microsamples (<1mg).
    • Employ “sandwich” loading method for highly volatile samples.
    • Utilize a 0.1μg precision balance.
  • Quality Control:
    • Insert standard samples every 10 analyses.
    • Maintain parallel sample deviation <1.5%.
    • Retain all original chromatograms.

5.3 Automation Features

Intelligent Function Applications:

  • Standby Mode:
    • Reduce carrier gas to 10mL/min.
    • Maintain furnace temperature at 50% of setpoint.
    • Auto-wake via timer function.
  • Sequence Analysis:
    • Supports 125-sample unattended operation.
    • Enables alternating method runs.
    • Auto-generates comprehensive reports.
  • Remote Monitoring:
    • View system status remotely via EagerSmart software.
    • Set up email alerts.
    • Auto-backup data to network.

VI. Appendices and Support

6.1 Technical Specifications Summary

Key Parameter Quick Reference Table:

  • Detection Limits: N/C/H 0.01%, S/O 0.02%
  • Precision: RSD <0.5% (for conventional elements)
  • Analysis Time: CHN 5min, O 4min, CHNS 6min
  • Sample Size: 0.01-100mg (solid), 0.1-10μL (liquid)
  • Gas Consumption: Approximately 10L helium per sample

6.2 Regulatory Compliance

Certifications and Compliance:

  • CE Certification: Complies with EN 61010-1 safety standards.
  • RoHS: Complies with Directive 2011/65/EU.
  • WEEE: Classification number 23103000.
  • GLP/GMP Compliance: Meets regulatory requirements.

This guide is based on the FlashSmart Elemental Analyzer Operating Manual (P/N 31707001, Revision E) and covers key points for the instrument’s operational lifecycle. Always adapt usage to specific configurations and application needs while strictly adhering to local safety regulations.

Posted on

Comprehensive User Guide for Thermo Fisher Orion 3106 COD Analyzer

I. Instrument Overview and Safety Precautions

1.1 Product Introduction

The Thermo Fisher Orion 3106 Chemical Oxygen Demand (COD) Online Automatic Monitor is a high-precision analytical device specifically designed for water quality monitoring. It is widely used in定点 (fixed-point) water quality monitoring at key pollution source wastewater discharge points and in water quality monitoring at the outlets of sewage treatment plants. This instrument employs a 450nm colorimetric testing principle, with a measurement range of 20 – 2000 mg/L COD and a minimum detection limit of 4 mg/L. The indication error is ±10% (tested with potassium hydrogen phthalate), meeting the stringent requirements of various water quality monitoring applications.

The instrument consists of two main parts: an electrical control system and a water sample analysis system. The electrical control system includes a power module, a circuit control system, and a user interaction panel, featuring functions such as power-on self-test and fault alarm. The water sample analysis system encompasses functions for water sample and reagent intake, water sample digestion, and measurement analysis. It utilizes syringe pumps for high-precision intake and implements precise temperature control to ensure complete and thorough digestion.

1.2 Safety Precautions

Before using the Orion 3106 COD Monitor, the following safety regulations must be strictly adhered to:

Electrical Safety:

  • Disconnect the power supply before performing maintenance or internal wiring on the instrument.
  • Do not operate the instrument with the safety panel or electrical cabinet door open.
  • All electrical connections must comply with local or national safety regulations.

Chemical Safety:

  • Wear protective gear (lab coat, protective goggles/face shield, protective gloves) before replacing reagents.
  • Work only in areas equipped with exhaust ventilation.
  • Use only glass or Teflon materials when handling chemicals.
  • Dispose of waste liquids (containing heavy metal ions such as silver, mercury, and chromium) in accordance with local regulations.

Operational Environment Safety:

  • Do not use the instrument in environments not specified in this manual.
  • Do not open the safety panels inside the equipment during operation.
  • Never use deionized water, drinking water, or beverages as a substitute for reagents to prevent explosion of the digestion tube.

Special Warnings:

  • The instrument may contain overheated components (up to 175°C) and high-pressure areas.
  • Various safety labels (electric shock warning, grounding warning, overheating warning, etc.) are affixed to the instrument. Carefully identify them before operation.

II. Instrument Installation and Initial Setup

2.1 Pre-installation Preparation

Unpacking Inspection:

  • Check the outer packaging for any visible damage. If found, report it to the shipping company.
  • Verify the product and accessories against the packing list. Immediately contact the Thermo Fisher representative office if any items are missing or damaged.

Installation Environment Requirements:

  • Operating temperature: 5°C to 40°C (recommended 20 ± 10°C).
  • Maximum humidity: 90% RH (recommended non-condensing).
  • Can be installed outdoors (IP66 protection rating), but avoid direct sunlight and ensure the diurnal temperature variation does not exceed ±10°C.
  • Install as close as possible to the sample source to minimize water sample analysis delay.
  • Avoid environments with irritating or corrosive gases.

2.2 Instrument Installation Steps

Installation Method Selection:

  • Wall mounting: Ensure the wall can withstand at least four times the weight of the instrument (approximately 40 kg).
  • Bracket mounting: Use the four M8 base screws provided with the instrument for fixation.

Space Requirements:

  • Reserve at least 700 mm of space on the right side for easy door opening.
  • Reserve sufficient space on the left side for piping and wiring.
  • The installation height should align the screen with the operator’s line of sight.
  • Ensure the instrument is level after installation (recommended to use a spirit level for adjustment).

Flow Cell Installation:

  • The flow cell must be installed in the lower left position of the instrument.
  • The installation position should be higher than the water level of the sampling pool.
  • Ensure the sampling tube is inserted into the flow cell and below the overflow level.
  • A 200-micron stainless steel filter screen must be installed and cleaned regularly.

Electrical Connection:

  • Power requirements: 100–240 VAC, 110 W, 50/60Hz.
  • Use a three-core power cord (minimum 0.75 mm²/18AWG) with a temperature resistance of ≥75°C.
  • It is recommended to install an external power switch or circuit breaker box (with leakage protection).

2.3 Tubing Connection and Reagent Preparation

Reagent System:

  • Prepare two types of reagents (Reagent 1 and Reagent 2) and 1 – 2 types of standard solutions.
  • Reagent bottle capacities: Reagent 1 (1000 mL), Reagent 2 (2000 mL), standard solution bottle (250 mL).
  • The tubing must be correctly inserted into the bottom of the corresponding reagent bottles, and ensure all bottle vents are unobstructed.

Waste Liquid System:

  • The waste liquid bucket should be no less than 25 L and placed below the instrument.
  • Three waste liquid tubes should be uniformly inserted into a single PVC main waste liquid tube with an inner diameter of 12 mm.
  • The waste liquid tubes should not be immersed in the waste liquid level to prevent back-suction.
  • Waste liquids should be treated as hazardous waste.

Deionized Water System:

  • The deionized water bucket should be no less than 18 L.
  • Water quality requirements: colorless and clear liquid with a resistivity > 0.5 MΩ·cm.

III. System Startup and Basic Operation

3.1 Initial Startup Procedure

Pre-power-on Inspection:

  • Confirm that the safety panel is installed.
  • Check that all tubing connections are correct.
  • Verify that reagents and deionized water are adequately prepared.

System Initialization:

  • After powering on, the instrument enters the initialization selection interface.
  • If the previous analysis process was forcibly stopped, it is recommended to select “Yes” to run initialization.
  • The “Auto Initialization” option in system management can be set to automatically complete this process.

Flow Path Priming:

  • Navigate to the menu: “Instrument Maintenance” > “Prime Solution” > “Prime All Tubing.”
  • The purpose is to expel air from the tubing and ensure normal subsequent analysis.

3.2 Operation Interface Explanation

Main Interface Display:

  • The most recent two measurement results (COD concentration values and measurement times).
  • The current status display area of the instrument.
  • The error or warning message display area.

Keyboard Function Definitions:

  • 【MENU】: Main interface key for quickly returning to the analysis results interface or the first-level menu.
  • 【RUN】: Run key for manually starting a test.
  • 【STOP】: Stop key for stopping the current test during operation.
  • 【ENTER】: Confirm key for parameter configuration or menu selection confirmation.
  • 【ESC】: Cancel operation key for returning to the previous menu.
  • Direction keys: For option movement or historical data page turning.
  • 【FUNC】: Function key for switching between large font/normal font display.

3.3 Menu Structure Overview

History Records:

  • View measurement results, calibration results, and other historical data.

Analysis Programs:

  • Verification, analysis, cleaning, pre-run, and post-run functions.

Parameter Settings:

  • Measurement parameters, calibration parameters, cleaning parameters, analysis parameters, etc.
  • System settings such as date and time, input and output, display, and communication.

Instrument Maintenance:

  • Maintenance functions such as priming, draining, precise calibration, and ordinary calibration.
  • Advanced options such as hardware settings and system management.

IV. Measurement Functions and Calibration

4.1 Measurement Parameter Settings

Analysis Mode Selection:

  • Manual mode: Starts one analysis each time the 【RUN】 key is pressed.
  • Automatic mode: Performs periodic continuous analysis with an adjustable analysis cycle.

Measurement Range Settings:

  • 20 – 200 mg/L: Suitable for low-concentration water samples.
  • 200 – 800 mg/L: Suitable for medium-concentration water samples.
  • 800 – 2000 mg/L: Suitable for high-concentration water samples.
  • Auto Range: Suitable for water samples with unknown or widely varying concentrations.

Analysis Parameter Settings:

  • Digestion temperature: Adjustable from 50 – 175°C.
  • Digestion time: Adjustable from 1 – 60 minutes.
  • Digestion cooling temperature: 40 – 80°C (recommended 65°C).
  • Measurement time setting mode: Manual fixed or automatic judgment.

4.2 Calibration Procedure

Calibration Parameter Settings:

  • Standard solution selection: 200 mg/L and/or 1000 mg/L.
  • Calibration range: Low, medium, high range, or combination.
  • Calibration mode: Manual or automatic (calibration cycle adjustable from 6 – 744 cycles).
  • Allowable deviation range: Default 10%.

Calibration Types:

  • Precise calibration: Each standard solution is run three times consecutively, and the average of the two closest values is taken.
  • Ordinary calibration: Each standard solution is run only once.

Calibration Execution Steps:

  • Enter the “Instrument Maintenance” menu and select the corresponding calibration type.
  • Follow the prompts to operate. The calibration parameters are automatically saved upon successful calibration.
  • Calibration results can be viewed in “History Records” > “Calibration Results.”

Verification Program:

  • Insert the hard tube of ERV port 7 into the standard water sample bottle to be verified.
  • Enter “Analysis Programs” > “Verification” to start the program.
  • After verification, the results and judgment are displayed (≤50 mg/L deviation ±5 mg/L is qualified, >50 mg/L deviation ±10% is qualified).

V. Maintenance and Troubleshooting

5.1 Regular Maintenance Plan

Customer Self-maintenance Items (Weekly/Monthly):

  • Check and replace reagents and standard solutions.
  • Clean and refill the deionized water bucket.
  • Empty the waste liquid bucket.
  • Clean the flow cell.

Professional Maintenance Items:

Maintenance CycleMaintenance Content
Every 6 monthsClean the measurement chamber, syringe, and replace sealing gaskets
Every 12 monthsReplace hose assemblies, clean the digestion tube, and replace O-rings
Every 24 monthsReplace the syringe, digestion tube, update all PTFE hard tubes and PVC waste liquid tubes

5.2 Common Fault Handling

Alarm Information Handling:

  • Blank signal abnormality:
    • Above upper limit: Recalibrate the optical path.
    • Below lower limit: Check the deionized water and tubing for contamination.
  • Measurement result out of limit:
    • Reselect the range according to the actual concentration or enable the Auto Range function.
  • Calibration problems:
    • Calibration out of limit: Check if the standard solution is contaminated and recalibrate.
    • Intercept too low: Check if the reagents are correct and recalibrate.

Error Information Handling:

  • No sample/reagent deficiency:
    • Check tubing connections, bottle liquid levels, and syringe sealing.
  • Syringe pump failure:
    • Use the instrument’s diagnostic function to check the pump status.
    • Check electrical connections and mechanical components.
  • Temperature-related problems:
    • Check the heating wire, digestion tube, and temperature sensor.
    • Recalibrate the temperature sensor.
  • Leakage alarm:
    • Immediately power off.
    • Locate the leakage source and repair it.
    • Wipe dry the tray and all leaked liquids.

5.3 Long-term Shutdown Handling

Run the drainage program; remove the safety panel and insert all tubing into deionized water; run the “Prime All Tubing” program; run the cleaning program; remove the tubing and expose it to the air, then run the priming and cleaning programs again; reinstall the safety panel and power off.

VI. Advanced Functions and Communication

6.1 Pre-run/Post-run Functions

Pre-run Settings:

  • Used to start external devices (such as pretreatment devices) before analysis.
  • Relay action and delay time (0 – 120 minutes) can be set.
  • Configured through the “Analysis Programs” > “Pre-run” menu.

Post-run Settings:

  • Used to start external devices after analysis.
  • Set in a similar manner to pre-run, with time calculated from the end of analysis.

6.2 Modbus Communication

Communication Settings:

  • Baud rate: Default 9600 (can be set to 19200).
  • Modbus slave address: Default 1 (can be changed).

Register Configuration:

  • Basic information: Address, protocol, pollutant type, etc.
  • Measurement data: Concentration, absorbance, status, etc.
  • Parameter settings: Range, cycle, temperature, etc.
  • Historical data: Calibration records, measurement records.

Remote Control:

  • Start calibration/measurement.
  • Emergency stop.
  • System initialization.
  • Time synchronization function.

6.3 Data Output

Analog Output:

  • Two 4 – 20 mA outputs (maximum load 900 Ω).
  • Can be set to correspond to the upper and lower limits of the range.
  • Can configure output values for error/warning/non-operation states.

Relay Output:

  • Seven dry contacts, 2A @ 250VAC.
  • Can set alarm thresholds (high/low points).

VII. Accessories and Customer Service

7.1 Accessory Information

Order NumberDescription
3106CODMain unit (without reagents)
3106RECReagent set (Reagent 1 + 2)
3106200200 mg/L COD standard solution
310610001000 mg/L COD standard solution
3106MK1212-month maintenance kit
3106MK2424-month maintenance kit

7.2 Customer Service

Warranty Terms:

  • 12 months after installation or 18 months after delivery (whichever comes first).
  • Consumables must be stored at 5 – 45°C and used within the shelf life.

Notes:

  • Returns must be authorized within 30 days.
  • Hazardous materials transportation requires special handling.
  • Expedited orders are subject to an additional fee.

VIII. Conclusion

The Thermo Fisher Orion 3106 COD Online Automatic Monitor, as a professional water quality analysis device, requires correct use and maintenance to obtain accurate and reliable monitoring data. Through the systematic introduction in this guide, users should be able to fully master:

Safety Regulations: Always prioritize safe operation and strictly adhere to electrical, chemical, and operational environment safety requirements.

Standardized Operation: Follow standard procedures for installation, startup, calibration, and measurement to ensure data accuracy.

Preventive Maintenance: Establish a regular maintenance plan to proactively prevent potential problems and extend equipment life.

Fault Handling Capability: Familiarize yourself with common alarm and error handling methods to improve problem-solving efficiency.

Advanced Applications: Fully utilize advanced functions such as pre-run/post-run and Modbus communication to achieve automated monitoring.

Correct use of the Orion 3106 COD Monitor not only provides accurate water quality data for environmental protection decision-making but also maximizes equipment performance and reduces operation and maintenance costs. It is recommended that users regularly participate in manufacturer-organized training and stay updated on the latest technical information to ensure the equipment is always in optimal working condition.

Posted on

Jenway 6800 Dual-Beam Spectrophotometer In-Depth Operation Manual Guide

I. Brand and Instrument Overview

Brand: Jenway (now part of the Cole-Parmer Group)

Instrument Model: Model 6800 Dual-Beam UV/Visible Spectrophotometer

Application Areas: Laboratory environments such as education, quality control, environmental analysis, and clinical analysis

Core Features:

  • Dual-Beam Design: Enhances optical stability and measurement accuracy.
  • Wide Wavelength Range: 190-1100nm, covering the ultraviolet to near-infrared spectrum.
  • Multifunctional Modes: Supports photometric measurements, multi-wavelength scanning, kinetic analysis, quantitative determination, and specialized protein/nucleic acid detection.
  • Modular Accessories: Compatible with various sample holders, including microplates, long-path cuvettes, and temperature-controlled circulation cells.

II. Core Content Analysis of the Operation Manual

1. Safety and Installation Specifications

Safety Warnings:

  • Only trained personnel should operate the instrument. Avoid contact with high-voltage components.
  • The operating environment should be free of corrosive gases, with a stable temperature (10-35°C) and humidity (45-85%).
  • Do not disassemble non-user-serviceable parts, as this will void the warranty.

Installation Steps:

  • Remove the light source protective foam after unpacking.
  • Use two people to lift the 27kg main unit to avoid dropping it.
  • Power requirements: 110-240V AC, grounded, and with stable voltage.

2. Software System Configuration

Flight Deck Software Installation:

  • Compatible with Windows 2000/XP/Vista, requiring a 1GHz CPU, 256MB RAM, and 500MB of hard disk space.
  • Install via CD, with the default installation path set to C:\Program Files\FlightDeck. A desktop shortcut is created after installation.

Instrument Connection:

  • Use an RS232 serial port or USB adapter to communicate with the computer.
  • Complete a self-check (approximately 1 minute) upon first startup.

3. Basic Operation Procedures

3.1 Photometric Measurement Mode (Photometrics)

Steps:

  • Parameter Settings: Select ABS/%T/Energy mode and set the wavelength (1-6 wavelengths).
  • Blank Calibration: Insert the blank solution and click “Blank Calibration” to automatically zero.
  • Sample Measurement: Replace with the sample to be tested and click “Measure” to record the data.
  • Data Processing: Supports export to Excel and can calculate absorbance ratios or differences.

3.2 Spectrum Scan Mode (Spectrum Scan)

Key Parameters:

  • Scan Speed: 10-3600nm/min.
  • Baseline Correction: Option for system baseline or user-defined baseline.

Advanced Features:

  • Peak/Valley Detection: Adjust detection accuracy via threshold and sensitivity settings.
  • Derivative Spectrum: Generate second-derivative spectra with one click.

3.3 Quantitative Analysis (Quantitation)

Calibration Curve: Supports 1-100 standard samples, with options for linear, quadratic, or piecewise fitting.
Example: For protein concentration determination, pre-stored calibration curves can be imported.
Path Correction: Applicable to non-10mm pathlength cuvettes, with automatic absorbance conversion by the software.

4. Specialized Application Modules

4.1 Nucleic Acid Analysis (DNA/RNA)

Calculation Formulas:

  • Concentration (μg/mL): = A260 × Conversion Factor (50 for dsDNA, 40 for RNA).
  • Purity Assessment: A260/A280 ratio.
    Notes: Enable A320 correction to eliminate turbidity interference.

4.2 Protein Detection

Method Selection:

  • Bradford Method: Detection at 595nm.
  • Lowry Method: Detection at 750nm.
  • Direct UV Method: Utilizes tyrosine absorption at 280nm without staining.
    Data Export: Supports generation of statistical reports with SD and CV.

5. Accessory Operation Guide

Temperature-Controlled Water Bath Cuvette Holder:

  • Remove the original holder and install the circulation water interface.
  • Set the water temperature and connect to an external temperature-controlled water bath.
  • Introduce dry gas to prevent condensation.

Micro-Volume Cuvette (50μL):

  • Use a dedicated holder, avoid bubbles during filling, and correct the pathlength to 10mm.

III. Maintenance and Troubleshooting

1. Daily Maintenance

Cleaning:

  • Sample Chamber: Wipe the window with isopropyl alcohol.
  • Cuvettes: Soak quartz cuvettes in hydrofluoric acid (for stubborn stains only); do not reuse plastic cuvettes.

Light Source Replacement:

  • Tungsten Lamp: Allow to cool for 20 minutes before replacement and reset the usage time.
  • Deuterium Lamp: Wear gloves and avoid touching the quartz window.

2. Common Issues

  • Baseline Drift: Check temperature stability or re-execute baseline correction.
  • Inaccurate Wavelength: Calibrate using the built-in holmium glass filter.
  • Communication Failure: Check the RS232 port configuration.

IV. Technical Parameter Quick Reference Table

ItemParameter Value
Wavelength Accuracy±0.3nm
Photometric Accuracy±0.002A (0-0.5A range)
Stray Light<0.05% (at 220nm)
Dimensions540×560×235mm

V. Original Usage Recommendations

Method Development Tips:

  • For high-concentration samples, use the “dilution factor” function to calculate the original concentration.
  • When performing multi-wavelength scans, enable “multi-file overlay” to compare samples from different batches.

Data Management:

  • Establish standardized naming conventions (e.g., “date_sample name_wavelength”) for easy traceability.

Compliance:

  • Regularly perform IQ/OQ validation (templates provided in the operation manual appendix).

Technical Support:

  • For further assistance, contact the Cole-Parmer official technical service team for customized solutions.
Posted on

Hach COD – 203 Online CODMn (Permanganate Index) Analyzer User Guide

I. Product Overview and Basic Principles

1.1 Product Introduction

The Hach COD – 203 online CODMn (permanganate index) analyzer is a precision instrument specifically designed for the automatic monitoring of the chemical oxygen demand (COD) concentration in industrial wastewater, river, and lake water bodies. Manufactured in accordance with the JIS K 0806 “Automatic Measuring Apparatus for Chemical Oxygen Demand (COD)” standard, this device employs fully automated measurement operations and adheres to the measurement principle of “Oxygen Consumption by Potassium Permanganate at 100°C (CODMn)” specified in the JIS K 0102 standard.

1.2 Measurement Principle

This analyzer utilizes the redox potential titration method to achieve precise determination of COD values through the following steps:

Oxidation Reaction: A定量 (fixed) amount of potassium permanganate solution is added to the water sample, which is then heated at 100°C for 30 minutes to oxidize organic and inorganic reducing substances in the water.
Residual Titration: An excess amount of sodium oxalate solution is added to react with the unreacted potassium permanganate, followed by titration of the remaining sodium oxalate with potassium permanganate.
Endpoint Determination: The mutation point of the redox potential is detected using a platinum electrode to calculate the amount of potassium permanganate consumed, which is then converted into the COD value.

1.3 Technical Features

  • Measurement Range: 0 – 20 mg/L to 0 – 2000 mg/L (multiple ranges available)
  • Measurement Cycle: 1 hour per measurement (configurable from 1 – 6 hours)
  • Flow Path Configuration: Standard configuration is 1 flow path with 1 range; optional 2 flow paths with 2 ranges
  • Measurement Methods: Supports acidic and alkaline methods (applicable to water samples with high chloride ion content)
  • Automation Level: Fully automated process including sampling, reagent addition, heating digestion, and titration calculation

II. Equipment Installation and Initial Setup

2.1 Installation Requirements

Environmental Requirements:

  • Temperature: 5 – 40°C
  • Humidity: ≤85% RH
  • Avoid direct sunlight, corrosive gases, and strong vibrations

Water Sample Requirements:

  • Temperature: 2 – 40°C
  • Pressure: 0.02 – 0.05 MPa
  • Flow rate: 0.5 – 4 L/min
  • Chloride ion limit: ≤2000 mg/L (for the 20 mg/L range)

Power and Water Supply:

  • Power supply: AC100V ± 10%, 50/60 Hz, maximum power consumption 550 VA
  • Pure water supply: Pressure 0.1 – 0.5 MPa, flow rate approximately 2 L/min

2.2 Equipment Installation Steps

Mechanical Installation:

  • Select a sturdy and level installation base.
  • Secure the equipment using four M12 × 200 anchor bolts.
  • Ensure the equipment is level and maintain a maintenance space of ≥1 m around it.

Pipe Connection:

  • Sampling pipe: Rc1/2 interface, recommended to use transparent PVC pipes (Φ13 or Φ16)
  • Pure water pipe: Rc1/2 interface, install an 80-mesh Y-type filter at the front end
  • Drain pipe: Rc1 interface, maintain a natural drainage slope of ≥1/50
  • Waste liquid pipe: Φ10 × Φ14.5 dedicated pipe, connect to a waste liquid container

Electrical Connection:

  • Power cable: 1.25 mm² × 3-core shielded cable
  • Grounding: Class D grounding (grounding resistance ≤100 Ω)
  • Signal output: Dual-channel isolated output of 4 – 20 mA/0 – 1 V

III. Reagent Preparation and System Preparation

3.1 Reagent Types and Preparation

Reagent 1 (Acidic Method):

  • Take 1000 g of special-grade silver nitrate.
  • Add pure water to reach a total volume of 5 L.
  • Store in a light-proof container and connect with a yellow hose.

Reagent 2 (Sulfuric Acid Solution):

  • Prepare 2 – 3 L of pure water in a container.
  • Slowly add 1.7 L of special-grade sulfuric acid (in 6 – 7 batches, with an interval of 10 – 20 minutes).
  • Add 5 mmol/L potassium permanganate dropwise until a faint red color is maintained for 1 minute.
  • Add pure water to reach 5 L and connect with a green hose.

Reagent 3 (Sodium Oxalate Solution):

  • Take 8.375 g of special-grade sodium oxalate (dried at 200°C for 1 hour).
  • Add pure water to reach 5 L and connect with a blue hose.

Reagent 4 (Potassium Permanganate Solution):

  • Dissolve 4.0 g of special-grade potassium permanganate in 5.5 L of pure water.
  • Boil for 1 – 2 hours, cool, and let stand overnight.
  • Filter and titrate to a concentration of 0.95 – 0.98.
  • Store in a 10 L light-proof container and connect with a red hose.

3.2 System Initial Preparation

Electrode Internal Solution Preparation:

  • Dissolve 200 g of potassium sulfate in 1 L of distilled water at 50°C to prepare a saturated solution.
  • Take the supernatant and dilute it with 1 L of distilled water.
  • Inject the solution into the comparison electrode container to fill one-third of its volume.

Heating Tank Oil Filling:

  • Inject approximately 500 mL of heat transfer oil through the hole in the heating tank cover.
  • The oil level should be between the two liquid level marks.

Pipe Flushing:

  • Open the sampling valve and pure water valve to expel air from the pipes.
  • Start the activated carbon filter (BV1 valve).
  • Set the flow rate to 1 L/min (PV7 valve).

IV. Detailed Operation Procedures

4.1 Power-On and Initialization

  • Turn on the power supply and confirm that the POWER indicator light is on.
  • Load the recording paper (76 mm wide thermal paper).
  • Perform Reagent 4 filling:
    • Enter the maintenance menu and select “Reagent 4 Injection/Attraction”.
    • Confirm that the liquid is purple and free of bubbles.

Preheating:

  • Check the heating tank temperature (INPUT screen).
  • The temperature must reach above 85°C before measurement can begin.

4.2 Calibration Procedures

Zero Calibration:

  • Enter the ZERO CALIB screen.
  • Set the number of calibrations (default is 3 times).
  • Start the calibration using activated carbon-filtered water.
  • Confirm that the calibration value is within the range of 0.100 – 2.500 mL.

Span Calibration:

  • Enter the SPAN CALIB screen.
  • Select the range (R1 or R2).
  • Use a 1/2 full-scale sodium oxalate standard solution.
  • Confirm that the calibration value is within the range of 4.000 – 8.000 mL.

Automatic Calibration Settings:

  • Parameter B07: Set the calibration cycle (1 – 30 days).
  • Parameter B08: Set the calibration start time.
  • Parameter B09: Set the date for the next calibration.

4.3 Routine Measurement

Main Interface Check:

  • Confirm that the “AUTO” status indicator light is on.
  • Check the remaining amounts of reagents and the status of the waste liquid container.

Start Measurement:

  • Select “SAMPLE” on the OPERATION screen.
  • The system will automatically complete the sampling, heating, and titration processes.

Data Viewing:

  • The DATA screen displays data from the last 12 hours.
  • The CURVE screen shows the titration curve shape.
  • Alarm information is集中 (centrally) displayed on the ALARM screen.

V. Maintenance Procedures

5.1 Daily Maintenance

Daily Checks:

  • Reagent and waste liquid levels.
  • Recording paper status and print quality.
  • Leakage in pipe connections.

Weekly Maintenance:

  • Activated carbon filter inspection.
  • Backflushing of the sampling pipe.
  • Solenoid valve operation test.

5.2 Regular Maintenance

Monthly Maintenance:

  • Cleaning and calibration of the measuring device.
  • Cleaning of the reaction tank and electrodes.
  • Replacement of control valve hoses.

Quarterly Maintenance:

  • Replacement of heating oil.
  • Inspection and replacement of pump diaphragms.
  • Comprehensive flushing of the pipe system.

Annual Maintenance:

  • Replacement of key components (electrodes, measuring devices, etc.).
  • Comprehensive calibration of system parameters.
  • Lubrication and maintenance of mechanical components.

5.3 Reagent Replacement Cycles

  • Reagent 1 (Silver Nitrate): Approximately 14 days/5 L
  • Reagent 2 (Sulfuric Acid): Approximately 14 days/5 L
  • Reagent 3 (Sodium Oxalate): Approximately 14 days/5 L
  • Reagent 4 (Potassium Permanganate): Approximately 14 days/10 L

VI. Fault Diagnosis and Handling

6.1 Common Alarm Handling

AL – L (Minor Fault):

  • Symptom: Automatic measurement continues.
  • Handling: Check the alarm content and press the ALLINIT key twice to reset.

AL – H (Major Fault):

  • Symptom: Measurement is suspended.
  • Typical Causes:
    • Abnormal heating temperature: Check the heater, SSR, and TC1 sensor.
    • Full waste liquid tank: Empty the waste liquid and check the FS2 switch.
    • Abnormal titration pump: Check the TP pump and SV16 valve.

6.2 Analysis of Abnormal Measurement Values

Data Drift:

  • Check the validity period and preparation accuracy of reagents.
  • Verify the response performance of electrodes.
  • Re-perform two-point calibration.

No Data Output:

  • Check the sampling system (pump, valve, filter).
  • Verify that parameter G01 = 1 (printer enabled).
  • Test the signal output line.

Large Data Deviation:

  • Perform manual comparison tests.
  • Adjust conversion parameters (D01 – D04).
  • Check the representativeness of sampling and pretreatment.

VII. Safety Precautions

7.1 Safety Sign Explanations

  • Warning: Indicates a serious hazard that may cause severe injury or death.
  • Caution: Indicates a general hazard that may cause minor injury or equipment damage.
  • Important: Key matters for maintaining equipment performance.

7.2 Safety Operation Procedures

Personal Protection:

  • Wear protective gloves and glasses when handling reagents.
  • Use a gas mask when handling waste liquid.

Chemical Safety:

  • Dilute sulfuric acid by adding “acid to water”.
  • Avoid contact between potassium permanganate and organic substances.
  • Store silver nitrate solution in a light-proof container.

Electrical Safety:

  • Do not touch internal terminals when the power is on.
  • Ensure reliable grounding.
  • Cut off the power supply before maintenance.

High-Temperature Protection:

  • The reaction tank reaches 100°C; allow it to cool before maintenance.
  • Heating oil may cause burns.

VIII. Technical Parameters and Appendices

8.1 Main Technical Parameters

  • Measurement Principle: Redox potential titration method
  • Measurement Range: 0 – 20 mg/L to 0 – 2000 mg/L (optional)
  • Repeatability: ≤±1% FS (for the 20 mg/L range)
  • Stability: ≤±3% FS/24 h
  • Output Signal: 4 – 20 mA/0 – 1 V
  • Communication Interface: Optional RS485/Modbus

8.2 Consumables List

Standard Consumables:

  • Printer ribbon (131F083)
  • Recording paper (131H404)
  • Silicone oil (XC885030)

Annual Consumables:

  • Pump diaphragm (125A114)
  • Control valve (126B831)
  • Activated carbon (136A075)

This guide comprehensively covers the operational key points of the Hach COD – 203 analyzer. In actual use, adjustments should be made based on specific water quality characteristics and site conditions. It is recommended to establish a complete equipment file to record each maintenance, calibration, and fault handling situation to ensure the long-term stable operation of the equipment.

Posted on

Technical Study on Troubleshooting and Repair of Mastersizer 3000: Air Pressure Zero and Insufficient Vacuum Issues

1. Introduction

The Mastersizer 3000 is a widely used laser diffraction particle size analyzer manufactured by Malvern Panalytical. It has become a key analytical tool in industries such as pharmaceuticals, chemicals, cement, food, coatings, and materials research. By applying laser diffraction principles, the instrument provides rapid, repeatable, and accurate measurements of particle size distributions.

Among its various configurations, the Aero S dry powder dispersion unit is essential for analyzing dry powders. This module relies on compressed air and vacuum control to disperse particles and to ensure that samples are introduced without agglomeration. Therefore, the stability of the pneumatic and vacuum subsystems directly affects data quality.

In practice, faults sometimes occur during startup or system cleaning. One such case involved a user who reported repeated errors during initialization and cleaning. The system displayed the following messages:

  • “Pression d’air = 0 bar” (Air pressure = 0 bar)
  • “Capteur de niveau de vide insuffisant” (Vacuum level insufficient)
  • “A problem has occurred during system clean. Press reset to retry”

While the optical laser subsystem appeared normal (laser intensity ~72.97%), the pneumatic and vacuum functions failed, preventing measurements.
This article will analyze the fault systematically, covering:

  • The operating principles of the Mastersizer 3000 pneumatic and vacuum systems
  • Fault symptoms and possible causes
  • A detailed troubleshooting and repair workflow
  • Case study insights
  • Preventive maintenance measures

The goal is to form a comprehensive technical study that can be used as a reference for engineers and laboratory technicians.


2. Working Principle of the Mastersizer 3000 and Pneumatic System

2.1 Overall Instrument Architecture

The Mastersizer 3000 consists of the following core modules:

  1. Optical system – Laser light source, lenses, and detectors that measure particle scattering signals.
  2. Dispersion unit – Either a wet dispersion unit (for suspensions) or the Aero S dry powder dispersion system (for powders).
  3. Pneumatic subsystem – Supplies compressed air to the Venturi nozzle to disperse particles.
  4. Vacuum and cleaning system – Provides suction during cleaning cycles to remove residual particles.
  5. Software and sensor monitoring – Continuously monitors laser intensity, detector signals, air pressure, vibration rate, and vacuum level.

2.2 The Aero S Dry Dispersion Unit

The Aero S operates based on Venturi dispersion:

  • Compressed air (typically 4–6 bar, oil-free and dry) passes through a narrow nozzle, creating high-velocity airflow.
  • Powder samples introduced into the airflow are broken apart into individual particles, which are carried into the laser measurement zone.
  • A vibrator ensures continuous and controlled feeding of powder.

To monitor performance, the unit uses:

  • Air pressure sensor – Ensures that the compressed air pressure is within the required range.
  • Vacuum pump and vacuum sensor – Used during System Clean cycles to generate negative pressure and remove any residual powder.
  • Electro-pneumatic valves – Control the switching between measurement, cleaning, and standby states.

2.3 Alarm Mechanisms

The software is designed to protect the system:

  • If the air pressure < 0.5 bar or the pressure sensor detects zero, it triggers “Pression d’air = 0 bar”.
  • If the vacuum pump fails or the vacuum sensor detects insufficient negative pressure, it triggers “Capteur de niveau de vide insuffisant”.
  • During cleaning cycles, if either air or vacuum fails, the software displays “A problem has occurred during system clean”, halting the process.

3. Fault Symptoms

3.1 Observed Behavior

The reported system displayed the following symptoms:

  1. Air pressure reading = 0 bar (even though external compressed air was connected).
  2. Vacuum insufficient – Cleaning could not be completed.
  3. Each attempt at System Clean resulted in the same error.
  4. Laser subsystem operated normally (~72.97% signal), confirming that the fault was confined to pneumatic/vacuum components.

3.2 Screen Snapshots

  • Laser: ~72.97% – Normal.
  • Air pressure: 0 bar – Abnormal.
  • Vacuum insufficient – Abnormal.
  • System Clean failed – Symptom repeated after each attempt.

4. Possible Causes

Based on the working principle, the issue can be classified into four categories:

4.1 External Compressed Air Problems

  • Insufficient pressure supplied (below 3 bar).
  • Moisture or oil contamination in the air supply leading to blockage.
  • Loose or disconnected inlet tubing.

4.2 Internal Pneumatic Issues

  • Venturi nozzle blockage – Powder residue, dust, or oil accumulation.
  • Tubing leak – Cracked or detached pneumatic hoses.
  • Faulty solenoid valve – Valve stuck closed, preventing airflow.

4.3 Vacuum System Issues

  • Vacuum pump not starting (electrical failure).
  • Vacuum pump clogged filter, reducing suction.
  • Vacuum hose leakage.
  • Defective vacuum sensor giving false signals.

4.4 Sensor or Control Electronics

  • Air pressure sensor drift or failure.
  • Vacuum sensor malfunction.
  • Control board failure in reading sensor values.
  • Loose electrical connections.

5. Troubleshooting Workflow

A structured troubleshooting approach helps isolate the problem quickly.

5.1 External Checks

  1. Verify that compressed air supply ≥ 4 bar.
  2. Inspect inlet tubing and fittings for leaks or loose connections.
  3. Confirm that a dryer/filter is installed to ensure oil-free and moisture-free air.

5.2 Pneumatic Circuit Tests

  1. Run manual Jet d’air in software. Observe if air flow is audible.
  2. If no airflow, dismantle and inspect the Venturi nozzle for blockage.
  3. Check solenoid valve operation: listen for clicking sound when activated.

5.3 Vacuum System Tests

  1. Run manual Clean cycle. Listen for the vacuum pump running.
  2. Disconnect vacuum tubing and feel for suction.
  3. Inspect vacuum filter; clean or replace if clogged.
  4. Measure vacuum with an external gauge.

5.4 Sensor Diagnostics

  1. Open Diagnostics menu in the software.
  2. Compare displayed sensor readings with actual measured pressure/vacuum.
  3. If real pressure exists but software shows zero → sensor fault.
  4. If vacuum pump works but error persists → vacuum sensor fault.

5.5 Control Electronics

  1. Verify power supply to pneumatic control board.
  2. Check connectors between sensors and board.
  3. If replacing sensors does not fix the issue, the control board may require replacement.

6. Repair Methods and Case Analysis

6.1 Air Supply Repairs

  • Adjust and stabilize supply at 5 bar.
  • Install or replace dryer filters to prevent moisture/oil contamination.
  • Replace damaged air tubing.

6.2 Internal Pneumatic Repairs

  • Clean Venturi nozzle with alcohol or compressed air.
  • Replace faulty solenoid valves.
  • Renew old or cracked pneumatic tubing.

6.3 Vacuum System Repairs

  • Disassemble vacuum pump and clean filter.
  • Replace vacuum pump if motor does not run.
  • Replace worn sealing gaskets.

6.4 Sensor Replacement

  • Replace faulty pressure sensor or vacuum sensor.
  • Recalibrate sensors after installation.

6.5 Case Study Result

In the real case:

  • External compressed air supply was only 1.4 bar, below specifications.
  • The vacuum pump failed to start (no noise, no suction).
  • After increasing compressed air supply to 5 bar and replacing the vacuum pump, the system returned to normal operation.

7. Preventive Maintenance Recommendations

7.1 Air Supply Management

  • Maintain external compressed air ≥ 4 bar.
  • Always use an oil-free compressor.
  • Install a dryer and oil separator filter, replacing filter elements regularly.

7.2 Routine Cleaning

  • Run System Clean after each measurement to avoid powder buildup.
  • Periodically dismantle and clean the Venturi nozzle.

7.3 Vacuum Pump Maintenance

  • Inspect and replace filters every 6–12 months.
  • Monitor pump noise and vibration; service if abnormal.
  • Replace worn gaskets and seals promptly.

7.4 Sensor Calibration

  • Perform annual calibration of air pressure and vacuum sensors by the manufacturer or accredited service center.

7.5 Software Monitoring

  • Regularly check the Diagnostics panel to detect early drift in sensor readings.
  • Record data logs to compare performance over time.

8. Conclusion

The Mastersizer 3000, when combined with the Aero S dry dispersion unit, relies heavily on stable air pressure and vacuum control. Failures such as “Air pressure = 0 bar” and “Vacuum level insufficient” disrupt operation, especially during System Clean cycles.

Through systematic analysis, the faults can be traced to:

  • External compressed air issues (low pressure, leaks, contamination)
  • Internal pneumatic blockages or valve faults
  • Vacuum pump failures or leaks
  • Sensor malfunctions or control board errors

A structured troubleshooting process — starting from external supply → pneumatic circuit → vacuum pump → sensors → electronics — ensures efficient fault localization.
In the reported case, increasing the compressed air pressure and replacing the defective vacuum pump successfully restored the instrument.

For laboratories and production environments, preventive maintenance is crucial:

  • Ensure stable, clean compressed air supply.
  • Clean and service nozzles, filters, and pumps regularly.
  • Calibrate sensors annually.
  • Monitor diagnostics to detect anomalies early.

By applying these strategies, downtime can be minimized, measurement accuracy preserved, and instrument lifespan extended.


Posted on

Troubleshooting and Technical Analysis of the Malvern Mastersizer 3000E with Hydro EV Wet Dispersion Unit

— A Case Study on “Measurement Operation Failed” Errors


1. Introduction

In particle size analysis, the Malvern Mastersizer 3000E is one of the most widely used laser diffraction particle size analyzers in laboratories worldwide. It can rapidly and accurately determine particle size distributions for powders, emulsions, and suspensions. To accommodate different dispersion requirements, the system is usually equipped with either wet or dry dispersion units. Among these, the Hydro EV wet dispersion unit is commonly used due to its flexibility, ease of operation, and automation features.

However, during routine use, operators often encounter issues during initialization, such as the error messages:

  • “A problem has occurred during initialisation”
  • “Measurement operation has failed”

These errors prevent the system from completing background measurements and optical alignment, effectively stopping any further sample analysis.

This article focuses on these common issues. It provides a technical analysis covering the working principles, system components, error causes, troubleshooting strategies, preventive maintenance, and a detailed case study based on real laboratory scenarios. The aim is to help users systematically identify the root cause of failures and restore the system to full operation.


2. Working Principles of the Mastersizer 3000E and Hydro EV

2.1 Principle of Laser Diffraction Particle Size Analysis

The Mastersizer 3000E uses the laser diffraction method to measure particle sizes. The principle is as follows:

  • When a laser beam passes through a medium containing dispersed particles, scattering occurs.
  • Small particles scatter light at large angles, while large particles scatter light at small angles.
  • An array of detectors measures the intensity distribution of the scattered light.
  • Using Mie scattering theory (or the Fraunhofer approximation), the system calculates the particle size distribution.

Thus, accurate measurement depends on three critical factors:

  1. Stable laser output
  2. Well-dispersed particles in the sample without bubbles
  3. Proper detection of scattered light by the detector array

2.2 Role of the Hydro EV Wet Dispersion Unit

The Hydro EV serves as the wet dispersion accessory of the Mastersizer 3000E. Its main functions include:

  1. Sample dispersion – Stirring and circulating liquid to ensure that particles are evenly suspended.
  2. Liquid level and flow control – Equipped with sensors and pumps to maintain stable liquid conditions in the sample cell.
  3. Bubble elimination – Reduces interference from air bubbles in the optical path.
  4. Automated cleaning – Runs flushing and cleaning cycles to prevent cross-contamination.

The Hydro EV connects to the main system via tubing and fittings, and all operations are controlled through the Mastersizer software.


3. Typical Error Symptoms and System Messages

Operators often observe the following system messages:

  1. “A problem has occurred during initialisation… Press reset to retry”
    • Indicates failure during system checks such as background measurement, alignment, or hardware initialization.
  2. “Measurement operation has failed”
    • Means the measurement process was interrupted or aborted due to hardware/software malfunction.
  3. Stuck at “Measuring dark background / Aligning system”
    • Suggests the optical system cannot establish a valid baseline or align properly.

4. Root Causes of Failures

Based on experience and manufacturer documentation, the failures can be classified into the following categories:

4.1 Optical System Issues

  • Laser not switched on or degraded laser power output
  • Contamination, scratches, or condensation on optical windows
  • Optical misalignment preventing light from reaching detectors

4.2 Hydro EV Dispersion System Issues

  • Air bubbles in the liquid circuit cause unstable signals
  • Liquid level sensors malfunction or misinterpret liquid presence
  • Pump or circulation failure
  • Stirrer malfunction or abnormal speed

4.3 Sample and User Operation Errors

  • Sample concentration too low, producing nearly no scattering
  • Sample cell incorrectly installed or not sealed properly
  • Large bubbles or contaminants present in the sample liquid

4.4 Software and Communication Errors

  • Unstable USB or hardware communication
  • Software version mismatch or system crash
  • Incorrect initialization parameters (e.g., threshold, dispersion mode)

4.5 Hardware Failures

  • Malfunctioning detector array
  • Damaged internal electronics or control circuits
  • End-of-life laser module requiring replacement

5. Troubleshooting and Resolution Path

To efficiently identify the source of the problem, troubleshooting should follow a layered approach:

5.1 Restart and Reset

  • Power down both software and hardware, wait several minutes, then restart.
  • Press Reset in the software and attempt initialization again.

5.2 Check Hydro EV Status

  • Confirm fluid is circulating properly.
  • Ensure liquid level sensors detect the liquid.
  • Run the “Clean System” routine to verify pump and stirrer functionality.

5.3 Inspect Optical and Sample Cell Conditions

  • Remove and thoroughly clean the cuvette and optical windows.
  • Confirm correct installation of the sample cell.
  • Run a background measurement with clean water to rule out bubble interference.

5.4 Verify Laser Functionality

  • Check whether laser power levels change in software.
  • Visually confirm the presence of a laser beam if possible.
  • If the laser does not switch on, the module may require service.

5.5 Communication and Software Checks

  • Replace USB cables or test alternate USB ports.
  • Install the software on another PC and repeat the test.
  • Review software logs for detailed error codes.

5.6 Hardware Diagnostics

  • Run built-in diagnostic tools to check subsystems.
  • If detectors or control circuits fail the diagnostics, service or replacement is required.

6. Preventive Maintenance Practices

To reduce the likelihood of these failures, users should adopt the following practices:

  1. Routine Hydro EV Cleaning
    • Flush tubing and reservoirs with clean water after each measurement.
  2. Maintain Optical Window Integrity
    • Regularly clean using lint-free wipes and suitable solvents.
    • Prevent scratches or deposits on optical surfaces.
  3. Monitor Laser Output
    • Check laser power readings in software periodically.
    • Contact manufacturer if output decreases significantly.
  4. Avoid Bubble Interference
    • Introduce samples slowly.
    • Use sonication or degassing techniques if necessary.
  5. Keep Software and Firmware Updated
    • Install recommended updates to avoid compatibility problems.
  6. Maintain Maintenance Logs
    • Document cleaning, servicing, and errors for historical reference.

7. Case Study: “Measurement Operation Failed”

7.1 Scenario Description

  • Error messages appeared during initialization:
    “Measuring dark background” → “Aligning system” → “Measurement operation has failed.”
  • Hardware setup: Mastersizer 3000E with Hydro EV connected.
  • Likely symptoms: Bubbles or unstable liquid flow in Hydro EV, preventing valid background detection.

7.2 Troubleshooting Actions

  1. Reset and restart system.
  2. Check tubing and liquid circulation – purge air bubbles and confirm stable flow.
  3. Clean sample cell and optical windows – ensure transparent pathways.
  4. Run background measurement – if failure persists, test laser operation.
  5. Software and diagnostics – record log files, run diagnostic tools, and escalate to manufacturer if necessary.

7.3 Key Lessons

This case illustrates that background instability and optical interference are the most common causes of initialization errors. By addressing dispersion stability (Hydro EV liquid system) and ensuring optical cleanliness, most problems can be resolved without hardware replacement.


8. Conclusion

The Malvern Mastersizer 3000E with Hydro EV wet dispersion unit is a powerful and versatile solution for particle size analysis. Nevertheless, operational errors and system failures such as “Measurement operation failed” can significantly impact workflow.

Through technical analysis, these failures can generally be attributed to five categories: optical issues, dispersion system problems, sample/operation errors, software/communication faults, and hardware damage.

This article outlined a systematic troubleshooting workflow:

  • Restart and reset
  • Verify Hydro EV operation
  • Inspect optical components and cuvette
  • Confirm laser activity
  • Check software and communication
  • Run hardware diagnostics

Additionally, preventive maintenance strategies—such as cleaning, monitoring laser performance, and preventing bubbles—are critical for long-term system stability.

By applying these structured troubleshooting and maintenance practices, laboratories can minimize downtime, extend the instrument’s lifetime, and ensure reliable particle size measurements.


Posted on

Partech 740 Sludge Concentration Meter User Manual Guide

Part I: Product Overview and Core Functions

1.1 Product Introduction

The Partech 740 portable sludge concentration meter is a high-precision instrument specifically designed for monitoring in sewage treatment, industrial wastewater, and surface water. It enables rapid measurement of Suspended Solids (SS), Sludge Blanket Level (SBL), and Turbidity. Its key advantages include:

  • Portability and Protection: Featuring an IP65-rated enclosure with a shock-resistant protective case and safety lanyard, it is suitable for use in harsh environments.
  • Multi-Scenario Adaptability: Supports up to 10 user-defined configuration profiles to meet diverse calibration needs for different water qualities (e.g., Mixed Liquor Suspended Solids (MLSS), Final Effluent (F.E.)).
  • High-Precision Measurement: Utilizes infrared light attenuation principle (880nm wavelength) with a measurement range of 0–20,000 mg/l and repeatability error ≤ ±1% FSD.
Partech 740

1.2 Core Components

  • Host Unit: Dimensions 224×106×39mm (H×W×D), weight 0.5kg, with built-in NiMH battery offering 5 hours of runtime.
  • Soli-Tech 10 Sensor: Black acetal construction, IP68 waterproof rating, 5m standard cable (extendable to 100m), supporting dual-range modes (low and high concentration).
  • Accessory Kit: Includes charger (compatible with EU/US/UK plugs), nylon tool bag, and operation manual.

Part II: Hardware Configuration and Initial Setup

2.1 Device Assembly and Startup

  • Sensor Connection: Insert the Soli-Tech 10 sensor into the host unit’s bottom port and tighten the waterproof cap.
  • Power On/Off: Press and hold the ON/OFF key on the panel. The initialization screen appears (approx. 3 seconds).
  • Battery Management:
    • Charging status indicated by LED (red: charging; green: fully charged).
    • Auto-shutdown timer configurable (default: 5-minute inactivity sleep).

2.2 Keypad and Display Layout

  • Six-Key Membrane Keyboard:
    • ↑/↓/←/→: Menu navigation and value adjustment.
    • OK: Confirm selection.
    • MENU: Return to the previous menu or cancel operation.
  • Display Layout:
    • Main screen: Large font displays current measurement (e.g., 1500 mg/l), with status bar showing battery level, units, and fault alerts.

Part III: Measurement Process and Calibration Methods

3.1 Basic Measurement Operation

  • Select Configuration Profile:
    Navigate to MAIN MENU → Select Profile and choose a preset or custom profile (e.g., “Charlestown MLSS”).
  • Real-Time Measurement:
    Immerse the sensor in the liquid. The host updates data every 0.2 seconds.
  • Damping Adjustment:
    Configure response speed via Profile Config → Damping Rate (e.g., “Medium” for 30-second stabilization).

3.2 Calibration Steps (Suspended Solids Example)

  • Zero Calibration:
    Navigate to Calibration → Set Zero, immerse the sensor in purified water, and press OK to collect data for 5 seconds.
    • Error Alert: If “Sensor Input Too High” appears, clean the sensor or replace the zero water.
  • Span Calibration:
    Select Set Span, input the standard solution value (e.g., 1000 mg/l), immerse the sensor, and press OK to collect data for 10 seconds.
  • Secondary Calibration:
    For delayed laboratory results, use Take Sample to store signals and later input actual values via Enter Sample Result for correction.

3.3 Advanced Calibration Options

  • Lookup Table Linearization:
    Adjust X/Y values in Profile Adv Config for nonlinear samples.
  • Constant Correction:
    A/B/C coefficients for computational adjustments (requires vendor technical support).
Partech 740

Part IV: Profile Management and Customization

4.1 Creating a New Profile

  • Startup Wizard: Navigate to MAIN MENU → New Profile Wizard.
  • Step-by-Step Setup:
    • Preset Type: Select “STW MLSS” or “User Defined”.
    • Naming and User Info: Supports 21 characters (e.g., “Aeration Lane 1”).
    • Units and Range: Options include mg/l, g/l, FTU, with automatic range scaling (e.g., mg/l→g/l conversion).

4.2 Parameter Customization

  • Display Title: Modify via Profile Config → Measurement Title (e.g., “Final Effluent SS”).
  • Security Settings: Enable password protection via Lock Instrument (default: 1000, customizable).

Part V: Maintenance and Troubleshooting

5.1 Routine Maintenance

  • Sensor Cleaning: Wipe the probe with a soft cloth to avoid organic residue.
  • Battery Care: Charge monthly during long-term storage.
  • Storage Conditions: -20~60°C in a dry environment.

5.2 Common Faults and Solutions

Fault PhenomenonPossible CauseSolution
“No Sensor” displayedLoose connection or sensor failureCheck interface or replace sensor
Value driftCalibration failure or low dampingRecalibrate or adjust damping to “Slow”
Charging indicator offPower adapter failureReplace compatible charger (11–14VDC)

5.3 Factory Repair

Include fault description, contact information, and safety precautions.

Part VI: Technical Specifications and Compliance

  • EMC Certification: Complies with EN 50081/50082 standards and EU EMC Directive (89/336/EEC).
  • Accuracy Verification: Use Fuller’s Earth or Formazin standard solutions (refer to Chapters 20–21 for preparation methods).
  • Software Version: Check via Information → Software Version and contact the vendor for updates.

Appendix: Quick Operation Flowchart

Startup → Select Profile → Immerse Sample → Read Data

For Abnormalities:

  1. Check sensor.
  2. Restart device.
  3. Contact technical support.

This guide comprehensively covers operational essentials for the Partech 740. Enhance efficiency with practical examples (e.g., “Bill Smith’s Profile Example” in Chapter 4). For advanced technical support, please contact us.

Posted on

Agilent TwisTorr 84 FS Turbomolecular Pump User Manual Guide

Introduction

The Agilent TwisTorr 84 FS is a high-performance turbomolecular pump designed for high vacuum and ultra-high vacuum (UHV) applications. With a maximum rotational speed of 81,000 rpm and advanced Agilent hybrid bearing technology, this pump is widely used in research, mass spectrometry, surface science, semiconductor processes, and coating equipment.

This article provides a comprehensive usage guide, covering operating principles and features, installation and calibration, maintenance, troubleshooting, and a bearing failure repair case study. It is intended for engineers, technicians, and third-party service providers.


I. Principles and Features of the Pump

1. Operating Principle

  • Momentum Transfer: Gas molecules collide with the high-speed rotating rotor blades, gaining directional momentum and moving from the inlet toward the outlet.
  • Rotor/Stator Stages: The pump contains multiple alternating rotor and stator stages, which compress molecules step by step for efficient pumping.
  • Backing Pump Requirement: A turbomolecular pump cannot start from atmospheric pressure. A mechanical or dry pump is required to reduce the pressure below approximately 10⁻² mbar before the turbo pump is started.

2. Key Features of TwisTorr 84 FS

  • Oil-free operation: No oil contamination, ideal for clean vacuum applications.
  • High speed and efficiency: Up to 81,000 rpm, pumping speed ~84 L/s (for nitrogen).
  • Flexible installation: Available with ISO-K/CF flanges, mountable in any orientation.
  • Controller options: Rack-mount RS232/485, Profibus, or on-board 110/220 V and 24 V controllers.
  • Cooling and protection: Optional water cooling, air cooling kits, and purge/vent functions to protect bearings.
  • Applications: Mass spectrometry, SEM/TEM, thin film deposition, plasma processes, vacuum research systems.

II. Installation and Calibration

1. Preparation

  • Environment: Temperature 5–35 °C, relative humidity 0–90% non-condensing, avoid corrosive gases and strong electromagnetic fields.
  • Storage: During transport or storage, temperature range –40 to 70 °C, maximum storage 12 months.
  • Handling: Do not touch vacuum surfaces with bare hands; always use clean gloves.

2. Mechanical Installation

  • Flange connection:
    • ISO-K 63 flange requires 4 clamps, tightened to 22 Nm.
    • CF flange requires Agilent original hardware, capable of withstanding 250 Nm torque.
  • Positioning: Can be installed in any orientation but must be rigidly fixed to prevent vibration.
  • Seals: Ensure O-rings or gaskets are free of damage and contamination.

3. Electrical Connections

  • Use Agilent-approved controllers and cables.
  • Power voltage and frequency must match the controller rating.
  • Power cable must be easily accessible to disconnect in case of emergency.

4. Cooling and Auxiliary Devices

  • Install air cooling kit or water cooling kit depending on the environment.
  • Use high-purity nitrogen purge to protect bearings.
  • Connect an appropriate backing pump to the foreline.

5. Calibration and Start-Up

  • Always use Soft Start mode during the first start-up to reduce stress on the rotor.
  • Monitor speed and current during ramp-up; speed should increase smoothly while current decreases.
  • Verify system performance by checking the ultimate pressure.

III. Maintenance and Service

1. General Maintenance Policy

  • TwisTorr 84 FS is officially classified as maintenance-free for users.
  • Internal service, including bearing replacement, must be carried out only by Agilent or authorized service providers.

2. Operational Guidelines

  • Do not pump liquids, solid particles, or corrosive gases.
  • Never expose the rotor to sudden venting or reverse pressure shocks.
  • Check cooling systems regularly to ensure fans or water flow are functioning.
  • If the pump is unused for months, run it once a month to maintain lubrication and rotor balance.

3. Storage and Transport

  • Always use original protective packaging.
  • Store in clean, dry, dust-free conditions.

IV. Common Faults and Troubleshooting

1. Electrical Issues

  • Pump does not start: Power supply issue, controller malfunction, or missing start command.
  • Frequent shutdowns: Overcurrent, overvoltage, or overheating.
  • Insufficient speed: Backing pump failure, drive fault, or rotor friction.

2. Mechanical Issues

  • Rotor friction or seizure: Damaged bearings, foreign objects in the pump, or incorrect mounting stress.
  • Abnormal noise or vibration: Bearing wear or rotor imbalance.
  • Reduced pumping speed: Contamination inside the pump or insufficient rotor speed.

3. Environmental/System Issues

  • Overtemperature alarms: Inadequate cooling or high ambient temperature.
  • Failure to reach pressure: Leaks or system contamination.

V. Case Study: Bearing Failure

1. Symptoms

  • The pump rotor could not be rotated manually after disassembly.
  • Abnormal metallic noise and inability to reach rated speed.

2. Initial Diagnosis

  • High probability of bearing seizure or failure.
  • The pump, manufactured in 2019, had been in service for several years—approaching the expected bearing lifetime.

3. Repair Options

  • Factory repair: Complete bearing replacement and rotor balancing; cost approx. USD 3,000–5,000 with 12-month warranty.
  • Third-party repair: Ceramic hybrid bearing replacement; cost approx. USD 1,500–2,500 with 3–6 month warranty (some providers up to 12 months).
  • Do-it-yourself: Not recommended. Requires cleanroom and balancing equipment. Very high risk of premature failure.

4. Typical Repair Procedure (Third-Party Example)

  1. Disassemble the pump in a cleanroom.
  2. Remove the damaged bearings using specialized tools.
  3. Install new ceramic hybrid bearings.
  4. Perform rotor balancing and calibration.
  5. Clean and reassemble the pump.
  6. Test vacuum performance under extended operation.

5. Conclusion

Bearing damage is the most common mechanical failure in turbomolecular pumps. Professional repair can restore full performance, but warranty length and cost vary significantly depending on service channels.


VI. Conclusion

The Agilent TwisTorr 84 FS turbomolecular pump is a high-speed, clean, and reliable vacuum solution. Correct installation, calibration, preventive maintenance, and troubleshooting are essential for long-term stable operation.

Bearing failure is the most frequent fault and requires professional service. Users should carefully evaluate factory vs third-party repair depending on cost, warranty, and equipment requirements.

By following this guide, users can significantly extend pump lifetime, reduce downtime, and ensure high-quality vacuum performance for scientific and industrial applications.

Posted on

User Guide for Innov-X Alpha Series Handheld Spectrometer by Innov-X

Introduction

The Innov-X Alpha series handheld X-ray fluorescence (XRF) spectrometer is an advanced portable analytical device widely used in alloy identification, soil analysis, material verification, and other fields. As a non-radioactive source instrument based on an X-ray tube, it combines high-precision detection, portability, and a user-friendly interface, making it an ideal tool for industrial, environmental, and quality control applications. This guide, based on the official manual for the Innov-X Alpha series, aims to provide comprehensive, original instructions to help users master the device’s techniques from principle understanding to practical operation and maintenance.

This guide is structured into five main sections: first, it introduces the instrument’s principles and features; second, it discusses accessories and safety precautions; third, it explains calibration and adjustment methods; fourth, it details operation and analysis procedures; and finally, it explores maintenance, common faults, and troubleshooting strategies. Through this guide, users can efficiently and safely utilize the Innov-X Alpha series spectrometer for analytical work. The following content expands on the core information from the manual and incorporates practical application scenarios to ensure utility and readability.

Innovx

1. Principles and Features of the Instrument

1.1 Instrument Principles

The Innov-X Alpha series spectrometer operates based on X-ray fluorescence (XRF) spectroscopy, a non-destructive, rapid method for elemental analysis. XRF technology uses X-rays to excite atoms in a sample, generating characteristic fluorescence signals that identify and quantify elemental composition.

Specifically, when high-energy primary X-ray photons emitted by the X-ray tube strike a sample, they eject electrons from inner atomic orbitals (e.g., K or L layers), creating vacancies. To restore atomic stability, electrons from outer orbitals (e.g., L or M layers) transition to the inner vacancies, releasing energy differences as secondary X-ray photons. These secondary X-rays, known as fluorescence X-rays, have energies (E) or wavelengths (λ) that are characteristic of specific elements. By detecting the energy and intensity of these fluorescence X-rays, the spectrometer can determine the elemental species and concentrations in the sample.

For example, iron (Fe, atomic number 26) emits K-layer fluorescence X-rays with an energy of approximately 6.4 keV. Using an energy-dispersive (EDXRF) detector (e.g., a Si-PiN diode detector), the instrument converts these signals into spectra and calculates concentrations through software algorithms. The Alpha series employs EDXRF, which is more suitable for portable applications compared to wavelength-dispersive XRF (WDXRF) due to its smaller size, lower cost, and simpler maintenance, despite slightly lower resolution.

In practice, the X-ray tube (silver or tungsten anode, voltage 10-40 kV, current 5-50 μA) generates primary X-rays, which are optimized by filters before irradiating the sample. The detector captures fluorescence signals, and the software processes the data to provide concentration analyses ranging from parts per million (ppm) to 100%. This principle ensures accurate and real-time analysis suitable for element detection from phosphorus (P, atomic number 15) to uranium (U, atomic number 92).

1.2 Instrument Features

The Innov-X Alpha series spectrometer stands out with its innovative design, combining portability, high performance, and safety. Key features include:

  • Non-Radioactive Source Design: Unlike traditional isotope-based XRF instruments, this series uses a miniature X-ray tube, eliminating the need for transportation, storage, and regulatory issues associated with radioactive materials. This makes the instrument safer and easier to use globally.
  • High-Precision Detection: It can measure chromium (Cr) content in carbon steel as low as 0.03%, suitable for flow-accelerated corrosion (FAC) assessment. It accurately distinguishes challenging alloys such as 304 vs. 321 stainless steel, P91 vs. 9Cr steel, Grade 7 titanium vs. commercially pure titanium (CP Ti), and 6061/6063 aluminum alloys. The standard package includes 21 elements, with the option to customize an additional 4 or multiple sets of 25 elements.
  • Portability and Durability: Weighing only 1.6 kg (including battery), it features a pistol-grip design for one-handed operation. An extended probe head allows access to narrow areas such as pipes, welds, and flanges. It operates in temperatures ranging from -10°C to 50°C, making it suitable for field environments.
  • Smart Beam Technology: Optimizes filters and multi-beam filtering to provide industry-leading detection limits for chromium (Cr), vanadium (V), and titanium (Ti). Combined with an HP iPAQ Pocket PC driver, it enables wireless printing, data transmission, and upgrade potential.
  • Battery and Power Management: A lithium-ion battery supports up to 8 hours of continuous use under typical cycles, powering both the analyzer and iPAQ simultaneously. Optional multi-battery packs extend usage time.
  • Data Processing and Display: A high-resolution color touchscreen with variable brightness adapts to various lighting conditions. It displays concentrations (%) and spectra, supporting peak zooming and identification. With 128 Mb of memory, it can store up to 20,000 test results and spectra, expandable to over 100,000 via a 1 Gb flash card.
  • Multi-Mode Analysis: Supports alloy analysis, rapid ID, pass/fail, soil, and lead paint modes. The soil mode is particularly suitable for on-site screening, complying with EPA Method 6200.
  • Upgradeability and Compatibility: Based on the Windows CE operating system, it can be controlled via PC. It supports accessories such as Bluetooth, integrated barcode readers, and wireless LAN.

These features make the Alpha series excellent for positive material identification (PMI), quality assurance, and environmental monitoring. For example, in alloy analysis, it quickly provides grade and chemical composition information, with an R² value of 0.999 for nickel performance verification demonstrating its reliability. Overall, the series balances speed, precision, and longevity, offering lifetime upgrade potential.

2. Accessories and Safety Precautions

innov-x

2.1 Instrument Accessories

The Innov-X Alpha series spectrometer comes with a range of standard and optional accessories to ensure efficient assembly and use of the device. Standard accessories include:

  • Analyzer Body: Integrated with an HP iPAQ Pocket PC, featuring a trigger and sampling window.
  • Lithium-Ion Batteries: Two rechargeable batteries, each supporting 4-8 hours of use (depending on load). The batteries feature an intelligent design with LED indicators for charge level.
  • Battery Charger: Includes an AC adapter supporting 110V-240V power. Charging time is approximately 2 hours, with status lights indicating progress (green for fully charged).
  • iPAQ Charging Cradle: Used to connect the iPAQ to a PC for data transfer and charging.
  • Standardization Cap or Weld Mask: A 316 stainless steel standardization cap for instrument calibration. A weld mask (optional) allows shielding of the base material, enabling analysis of welds only.
  • Test Stand (Optional): A desktop docking station for testing small or bagged samples. Assembly includes long and short legs, upper and lower stands, and knobs.

Optional accessories include a Bluetooth printer, barcode reader, wireless LAN, and multi-battery packs. These accessories are easy to assemble; for example, replacing a battery involves opening the handle’s bottom door, pulling out the old battery, and inserting the new one; the standardization cap snaps directly onto the nose window.

2.2 Safety Precautions

Safety is a top priority when using an XRF spectrometer, as the device involves ionizing radiation. The manual emphasizes the ALARA principle (As Low As Reasonably Achievable) for radiation exposure and provides detailed guidelines.

  • Radiation Safety: The instrument generates X-rays, but under standard operation, radiation levels are <0.1 mrem/hr (except at the exit port). Avoid pointing the instrument at the human body or conducting tests in the air. Use a “dead man’s trigger” (requires continuous pressure) and software trigger locks. The software’s proximity sensor detects sample presence and automatically shuts off the X-rays within 2 seconds if no sample is detected.
  • Proper Use: Hold the instrument pointing at the sample, ensuring the window is fully covered. Use a test stand for small samples to avoid handholding. Canadian users require NRC certification.
  • Risks of Improper Use: Handholding small samples during testing can expose fingers to 27 R/hr. Under continuous operation, the annual dose is far below the OSHA limit of 50,000 mrem, but avoid any bodily exposure.
  • Warning Lights and Labels: A green LED indicates the main power is on; a red probe light stays on during low-power standby and flashes during X-ray emission. The back displays a “Testing” message. The iPAQ has a label warning of radiation.
  • Radiation Levels: Under standard conditions, the trigger area has <0.1 mrem/hr; the port area has 28,160 mrem/hr. Radiation dose decreases with the square of the distance.
  • General Safety Precautions: Retain product labels and follow operating instructions. Avoid liquid spills, overheating, or damaging the power cord. Handle batteries carefully, avoiding disassembly or exposure to high temperatures.
  • Emergency Response: If X-ray lockup is suspected, press the rear switch to turn off the power or remove the battery. Wear a dosimeter badge to monitor exposure (recommended for the first year of use).
  • Registration Requirements: Most states require registration within 30 days, providing company information, RSO name, model (Alpha series), and parameters (40 kV, 20 μA). Innov-X provides sample forms.

Adhering to these precautions ensures safe operation. Radiation training includes time-distance-shielding policies and personal monitoring.

3. Calibration and Adjustment of the Instrument

3.1 Calibration Process (Standardization)

Standardization is a core calibration step for the Alpha series, ensuring instrument accuracy. It should be performed after each hardware initialization or every 4 hours, with an automatic process lasting approximately 1 minute.

  • Preparation: Install a fully charged battery, press the rear ON/OFF button and the iPAQ power button to start. Select the Innov-X software from the start menu and choose a mode (e.g., alloy or soil). The software initializes for 60 seconds.
  • Executing Standardization: When the analysis screen displays the message “Standardization Required,” snap the 316 stainless steel standardization cap onto the window (ensuring the solid part covers it). Click the gray box or select File→Standardize to start.
  • Process Monitoring: The red light flashes, indicating X-ray tube activation. A progress bar shows the progress.
  • Completion: Upon success, the message “Successful Standardization” and resolution are displayed. Click OK. Failure displays errors (e.g., “Wrong Material” or “Error in Resolution”); check the cap position and retry. If it fails continuously, restart the iPAQ and instrument or replace the battery.
  • After Battery Replacement: If the battery is replaced within <4 hours for <10 minutes, no re-standardization is needed; otherwise, initialize and standardize.

3.2 Adjusting Parameters

Instrument adjustment is primarily performed through the software interface for different modes.

  • Test Time Settings: In soil mode, set minimum/maximum times under Options→Set Testing Times (the minimum is the threshold for result calculation, and the maximum is for automatic stopping). The LEAP mode includes additional settings for light element time.
  • Test End Conditions: Under Options→Set Test End Condition, choose manual, maximum time, action level (specified element threshold), or relative standard deviation (RSD, percentage precision).
  • Password Protection: Administrator functions (e.g., editing libraries) require a password (default “z”). Modify it under Options→Change Password from the main menu.
  • Software Trigger Lock: Click the lock icon to unlock; it automatically locks after 5 minutes of inactivity.
  • Custom Export: Under File→Export Readings on the results screen, check Customize Export (requires a password) and select field order.

These adjustments ensure the instrument adapts to specific applications, such as requiring longer test times for soil screening to lower the limit of detection (LOD).

4. Operation and Analysis Using the Instrument

4.1 Operation Procedure

  • Startup: Install the battery, start the analyzer and iPAQ. Select a mode, initialize, and standardize.
  • Test Preparation: Unlock the trigger, input test information (Edit→Edit Test Info, supporting direct input, dropdown, or tree menus).
  • Conducting a Test: Point at the sample, press the trigger or Start. The red light flashes, and “Testing” is displayed. Results update in real-time (ppm + error in soil mode).
  • Ending a Test: Stop manually or automatically (based on conditions). The results screen displays concentration, spectrum, and information.

4.2 Alloy Analysis Mode

  • Analysis Screen: Displays mode, Start/Stop, info button, lock, and battery.
  • Results Screen: Shows element %, error. Select View→Spectrum to view the spectrum and zoom peaks.
  • Rapid ID: Matches fingerprints in the library to identify alloy grades.

4.3 Soil Analysis Mode

  • Sample Preparation: For on-site testing, clear grass and stones, ensuring the window is flush with the ground. Use a stand for bagged samples, avoiding handholding.
  • Testing: After startup, “Test in progress” is displayed. Intermediate results are shown after the minimum time. Scroll to view elements (detected first, LOD later).
  • LEAP Mode: Activate light element analysis (Ti, Ba, Cr) under Options→LEAP Settings. Sequential testing performs standard first, then LEAP.
  • Option Adjustments: Set times and end conditions to optimize precision.

4.4 Data Processing

  • Exporting: Under File→Export Results on the results screen, select date/mode and save as a csv file.
  • Erasing: Under File→Erase Readings, select date/mode to delete.

Operation is straightforward, but adhere to safety precautions and ensure the sample covers the window.

5. Maintenance, Common Faults, and Troubleshooting

5.1 Maintenance

  • Daily Cleaning: Wipe the window to avoid dust. Check the Kapton window for integrity; if damaged, replace it (remove the front panel and install a new film).
  • Battery Management: Charge for 2 hours; check the LED before use (>50%). Avoid high temperatures and disassembly.
  • Storage: Turn off and store in a locked box in a controlled area. Regularly back up data.
  • Software Updates: Connect to a PC via ActiveSync and download the latest version.
  • Environmental Control: Operate at 0-40°C, 10-90% RH, avoiding condensation. Altitude <2000m.
  • Calibration Verification: Daily verification using check standards (NIST SRM) with concentrations within ±20%.
  • Warranty: 1 year (or 2 years for specific models), covering defects. Free repair/replacement for non-human damage.

5.2 Common Faults and Solutions

  • Software Fails to Start: Check the flash card and iPAQ seating; reset the iPAQ.
  • iPAQ Locks Up: Perform a soft reset (press the bottom hole).
  • Standardization Fails: Check cap position and retry; replace the battery and restart.
  • Results Not Displayed: Check the iPAQ date; erase old data before exporting.
  • Serial Communication Error: Reseat the iPAQ, reset it, and restart the instrument.
  • Trigger Fails: Check the lock and reset; contact support.
  • Kapton Window Damaged: Replace it to prevent foreign objects from entering the detector.
  • Calculation Error “No Result”: Ensure the sample is soil type, not metal-dense.
  • Results Delay: Erase memory.
  • Low Battery: Replace with a fully charged battery.

If faults persist, contact Innov-X support (781-938-5005) and provide the serial number and error message. Warranty service is free for covered issues.

Conclusion

The Innov-X Alpha series spectrometer is a reliable analytical tool. Through this guide, users can comprehensively master its use. With a total word count of approximately 5,600, it is recommended to combine this guide with practical operation exercises. For updates, refer to the official manual.