Posted on

NEXTorr® Z 100 ND Float Pump User Guide

1. Overview and Principle

The NEXTorr® Z 100 ND Float Pump is a hybrid ultra-high vacuum pump that combines a Non-Evaporable Getter (NEG) with a Sputter Ion Pump (SIP). The NEG element efficiently removes active gases such as H₂, CO, CO₂, O₂, and H₂O, while the ion pump handles inert gases (such as Ar) and methane, also providing a current signal that can be used as a pressure indication. The Z100 features compact size, low power consumption, and minimal magnetic interference, making it ideal for scanning electron microscopes and other sensitive equipment.

NEG works by chemically absorbing and dissolving gas molecules at room temperature, but it must first be activated at high temperature (about 400–500 °C for 1 hour) to remove the passivation layer. After activation, NEG continuously pumps at room temperature with virtually no power consumption. The ion pump operates by ionizing residual gas molecules under high electric and magnetic fields. Positive ions are accelerated to strike the cathode and become trapped. The “ND” (Noble Diode) design improves the pumping of inert gases.

2. Applications

  • Ultra-high vacuum chambers in SEMs
  • Compact research equipment with space constraints
  • Systems sensitive to vibration and magnetic fields
  • Environments with a significant inert gas background

3. Installation and Commissioning

3.1 Mechanical Installation

  • Verify flange type and sealing surfaces are clean and free of scratches.
  • Use copper gaskets or O-rings, tighten with proper torque.
  • Avoid vacuum grease contamination, keep the pump inlet clean.
  • Install away from strong magnetic fields of the electron optics.

3.2 Electrical Connection

  • The ion pump requires a high-voltage power supply (typically 3–7 kV).
  • The NEG requires heater/temperature control for activation.
  • Ensure HV cables are securely locked and correct polarity is applied.

3.3 Initial Pump Down and Leak Check

  • Use a forepump/turbo system to reach ≤10⁻⁶ mbar before activation.
  • Perform helium leak detection to confirm no flange leakage.

3.4 NEG Activation

  • Heat NEG under vacuum to 400–500 °C for about 1 hour.
  • Monitor vacuum level and ion pump current during activation.
  • Cool down to room temperature before normal operation.

3.5 Ion Pump Startup

  • Once good vacuum is established and NEG is activated, gradually apply HV to start the ion pump.
  • Monitor ion current decreasing trend as an indication of pressure.

4. Operation and Maintenance

  • Use ion pump current as a proxy for chamber pressure.
  • For long-term shutdown, fill chamber with dry nitrogen to prevent contamination.
  • NEG can be reactivated several times but capacity will decrease gradually.
  • Avoid hydrocarbons or oil vapors entering the vacuum system.

5. Common Failures and Troubleshooting

  1. Slow pumping or cannot reach target pressure: Possible leaks, unactivated NEG, contamination, or poor conductance. → Leak check, re-activation, bakeout.
  2. High ion pump current: Possible leaks, discharges, or wiring errors. → Inspect sealing, reduce HV, check wiring.
  3. NEG performance decline: May be saturated or surface contaminated. → Re-activate or replace NEG.
  4. HV discharges: May be due to insufficient vacuum or insulation issues. → Reduce HV, re-pump, clean cables.
  5. Unstable readings: Ion current depends on gas composition. → Cross-check with independent gauges.

6. Integration with SEM

  • Minimize Ar contamination from sample preparation.
  • Control activation temperature within SEM chamber tolerance.
  • Use ion pump current as interlock for SEM HV supply.
  • Maintain strict cleanliness to prevent NEG contamination.

7. Safety Notes

  • Ion pump power supply is high voltage; always power down and discharge before servicing.
  • NEG activation involves high temperature; ensure insulation and thermal compatibility.
  • Follow SEM manufacturer’s operational and safety guidelines.

8. Conclusion

The NEXTorr® Z 100 ND Float Pump combines the fast pumping speed of NEG with the full gas spectrum coverage of an ion pump. Its compact design, low power consumption, and long lifetime make it ideal for SEM and UHV applications. Proper installation, activation, and regular maintenance are essential to ensure stable long-term performance.


Posted on

User Guide for the Toledo ND245 Electronic Weighing Instrument Manual

Introduction

In modern industrial and commercial sectors, precise weighing is crucial for ensuring product quality, production efficiency, and fair trade. As a global leader in weighing solutions, Mettler Toledo’s IND245 Electronic Weighing Instrument (Vehicle Scale Version) stands out with its advanced technology, reliable performance, and flexible application scenarios, making it an ideal choice for vehicle weighing, logistics management, and industrial weighing. Designed specifically for vehicle scales, it supports both analog and digital sensor inputs, catering to scenarios such as truck scales and lorry weighbridges, and handling complex weighing needs from small vehicles to heavy-duty trucks.

IND245

This guide, based on the Technical Manual for the Toledo ND245 Electronic Weighing Instrument (Vehicle Scale Version), aims to provide users with a comprehensive and practical operational reference. It will start with the instrument’s principles, features, and specifications, followed by step-by-step instructions on installation and maintenance, daily operation procedures, and parameter settings, concluding with discussions on common faults and their resolution strategies. Through this guide, users will not only be able to get started quickly but also optimize instrument performance for long-term stable operation. Whether you are a first-time user or an experienced engineer, this guide will help you maximize the potential of the IND245 for efficient and accurate weighing management.

The IND245 is designed with a focus on user-friendliness and high reliability. It adopts a modular structure and supports multiple communication interfaces, suitable for a variety of applications ranging from simple weighing to complex vehicle pairing. The manual emphasizes the involvement of professionals in commissioning to avoid safety hazards. This guide will strictly adhere to the principles outlined in the manual, providing original interpretations and expanded explanations to help users apply the IND245 flexibly in real-world environments.

Instrument Principles, Features, and Specifications

Working Principles

The IND245 Electronic Weighing Instrument operates based on precise signal acquisition, processing, and display technologies. It connects to weighing sensors (either analog or digital types) to convert mechanical force into electrical signals, which are then digitized by an A/D converter and ultimately displayed on an LCD screen as weighing results. The core components include the mainboard, A/D conversion module, microprocessor, and display keyboard.

For analog sensors, the IND245 supports sensors with a 350-ohm load resistance and is compatible with sensitivities of 2mV/V and 3mV/V without additional configuration. The sensor converts weight changes into millivolt-level voltage signals, with the mainboard providing a 10V excitation voltage. The A/D converter performs high-speed digital conversion at a sampling rate of 366Hz. The microprocessor applies digital filtering algorithms (such as low-pass filtering and steady-state detection) to eliminate noise, ensuring accuracy within 6000e (verification divisions).

For digital sensors (such as the SLC720 POWERCELL GDD), the instrument uses the RS-422/485 protocol, supporting up to 12 sensors connected via a 300-meter Homerun cable. Digital signals are transmitted directly, avoiding attenuation and interference inherent in analog transmission, thereby enhancing anti-interference capabilities and precision stability. The instrument incorporates a real-time clock (RTC) and an SD/MicroSD card for data backup and Alibi storage, ensuring tamper-proof transaction records.

The overall principle can be summarized as: Sensor → Signal Excitation/Acquisition → A/D Conversion → Digital Filtering/Processing → Display/Output. The vehicle scale version is specifically optimized for paired weighing functions, supporting inbound/outbound operations, automatically calculating net weight, and ensuring positive output through negative net weight correction, suitable for logistics scenarios.

IND245

Key Features

The IND245 stands out for its versatility and cost-effectiveness, with key features including:

  1. High Precision and Wide Range: Supports up to 50,000 display divisions with an accuracy of 6000e. Automatic zero tracking (AZM) and multi-range switching ensure accurate measurements from微量 (trace amounts) to heavy loads. Adjustable steady-state detection time (0.3-1 second) enables fast dynamic response, suitable for vehicles quickly mounting the scale.
  2. Flexible Sensor Compatibility: Seamlessly supports 8 analog sensors or 12 digital sensors. The digital version maintains signal integrity over long distances, reducing wiring costs.
  3. Rich Communication and Integration Options: Standard RS-232/422/485 interfaces support SICS protocol, continuous output, and CTPZ commands. Optional interfaces include USB, Ethernet, and DIO (2 inputs, 4 outputs), facilitating integration with PLCs, PCs, or printers. The vehicle scale version includes built-in preset points and a tare library, supporting 100 temporary and 200 permanent tare records.
  4. User-Friendly Interface: A 240×96 dot-matrix LCD display supports Chinese and English switching. The 25-key keyboard includes numeric/alphabetic input and navigation keys, with unique digital shortcuts for accelerated menu navigation. The system row displays DIO status and time, while the information input area supports ID/vehicle number entry.
  5. Data Security and Storage: Alibi memory stores 60,000 transaction records, which are non-deletable. 4000 transaction logs and SD card backup support data recovery. Parameter locking in certification mode prevents tampering.
  6. Vehicle Scale-Specific Functions: Supports paired/standard/simple weighing modes, with negative net weight correction automatically swapping gross/tare weights. Preset point functionality allows setting target weight thresholds with advance warning, improving operational efficiency.
  7. Strong Environmental Adaptability: Stainless steel enclosure (IP66 dust and water resistance version), operating temperature range of -10°C to 40°C, and humidity tolerance of 10% to 95%. A 100-240VAC wide voltage input makes it suitable for outdoor vehicle scales.

These features enable the IND245 to excel in vehicle scale applications, such as calculating net weights for vehicles entering and exiting logistics parks, reducing human errors, and increasing throughput.

Technical Specifications

The specifications of the IND245 are detailed in Table 1-1 of the manual. Below is a summary of key parameters presented in a table for easy comparison:

Parameter CategorySpecification Details
Form FactorStandard/Dustproof (IP66), all stainless steel 304L; Tabletop/wall-mounted/pole-mounted installation
Dimensions (L×W×D)230 mm × 165.3 mm × 146.4 mm
WeightAnalog version: 3.2 kg; Digital version: 3.5 kg
Power Supply100–240 VAC, 50/60 Hz; Analog version: 750 mA; Digital version: 500 mA
Display240 × 96 LCD dot-matrix screen, refresh rate of 10 times/second, maximum divisions of 50,000
Sensor SupportAnalog: 8×350Ω, 2/3 mV/V; Digital: 12×SLC720 (POWERCELL GDD), 300m cable
A/D Update RateAnalog: 366 Hz; Digital: 25 Hz
Keyboard25 keys (numeric/alphabetic/navigation/function keys)
Communication InterfacesStandard: RS-232/422/485; Options: USB, Ethernet, DIO (2 inputs, 4 outputs)
StorageAlibi: 60,000 records; Transaction logs: 4,000 records; Tare table: 100 temporary/200 permanent
Environmental ConditionsTemperature: -10°C to 40°C; Humidity: 10% to 95% (non-condensing)
CertificationsChina Accuracy Class III, 6000e; OIML/USA/Canada options available

These specifications ensure the IND245’s reliable operation in industrial environments, supporting diverse needs from static vehicle weighing to dynamic logistics. Users can choose between analog and digital versions based on specific applications, with the digital version being more suitable for long-distance, multi-sensor scenarios.

How to Install and Maintain the Instrument?

Installation Guide

The installation of the IND245 must be carried out by professional personnel to ensure safe grounding and avoid live plugging and unplugging. Chapter 2 of the manual provides a detailed description of the process from unpacking to lead sealing.

1. Unpacking and Preparation

  • Opening the Instrument: Use a flat-head screwdriver to loosen the six stainless steel spring clips on the front cover (Figure 2-1). For the dustproof version, carefully release the bottom spring clips to avoid damaging the seal.
  • Environmental Protection: Not suitable for hazardous areas as it is non-explosion-proof. The dustproof version is IP66-rated, suitable for water washing environments but should avoid high temperatures and corrosion.

2. Installation Methods

The IND245 supports tabletop, wall-mounted, and pole-mounted installations:

  • Tabletop Installation: Attach four rubber pads to the bottom for anti-slip (Figure 2-3).
  • Wall-Mounted Installation: Use two brackets and four M5 screws for fixation. Rotate the front cover 180° to exchange the power/sensor cable entries (analog versions require adjustment; digital versions do not; Figures 2-4 to 2-8).
  • Pole-Mounted Installation: Similar to wall-mounted installation, using dedicated brackets and ensuring the ability to withstand four times the instrument’s self-weight.

Installation Location: Avoid direct sunlight and vibration sources, and ensure the distance to sensors does not exceed specified lengths.

3. Cabling and Wiring

  • Magnetic Ring Installation: Thread each cable through a magnetic ring and loop it near the housing to prevent interference (Figure 2-10).
  • Standard/Sealed Connectors: Use standard connectors for standard versions (Figure 2-11); select appropriate rubber rings for sealing in dustproof versions (Table 2-1, Figures 2-12 to 2-13).
  • Cable Configuration: Standard versions have eight interfaces (power, DIO, USB, Ethernet, COM1/2, sensors; Table 2-2). Dustproof analog versions have six sealed sleeves (Figures 2-15, Table 2-3).
  • Mainboard Wiring: Analog sensors can be connected using 4-wire or 6-wire configurations (Figures 2-17 to 2-18); digital sensors are connected using POWERCELL (Figure 2-19). Connect the AC power supply (L/N/GND; Figures 2-6/2-7).
  • Optional Component Connection: COM1 RS-232 (Figure 2-23); second serial port/USB/DIO/Ethernet (Section 2.4.10).
  • Switch Settings: Set the SW1 metering switch to ON (certification mode); select DIO switches for passive/active mode (Figure 2-66).

4. Final Steps

  • SD/MicroSD Card Installation: Insert into the mainboard slot (Figures 2-67/2-68) for Alibi/backup purposes.
  • Range Label: Affix a label beside the display indicating capacity/e value (Figures 2-69/2-70).
  • Closing the Housing: Press down on the four corners crosswise until a “click” sound is heard (Section 2.10).
  • Lead Sealing: In certification mode, thread a sealing wire through and fix it (Figure 2-71).

After installation, perform a functional test to ensure no short circuits or leakage currents.

Maintenance Guide

Regular maintenance ensures the long-term stability of the instrument. Chapter 5 of the manual emphasizes the importance of professional servicing.

1. Daily Cleaning

  • Clean the housing with a neutral detergent and a soft cloth, avoiding industrial solvents. Do not spray water onto the keyboard or display to prevent damage from sharp objects. Regularly inspect and maintain records.

2. Software Upgrades

  • Supports online upgrades. After downloading new firmware, perform a master reset (SW1-2/4 ON, power on to confirm). Back up SD card data to avoid memory errors.

3. Routine Inspections

  • Professional personnel should perform calibration once a year, checking sensors, cables, and grounding. Verify accuracy and clean internal dust.

4. Service Support

  • Contact Mettler Toledo’s service department for support. After on-site installation, only regular calibration is required. Use original factory parts for replacements.

Maintenance Principles: Always cut off the power before operating and keep the instrument dry. While the expected lifespan is long, more frequent inspections may be necessary in harsh environments.

What Are the Operation Procedures and Parameter Settings for the Instrument?

Operation Procedures

The IND245 is designed for ease of use, with Chapter 3 of the manual providing detailed information on the keyboard and main window.

1. Keyboard and Interface

  • Keyboard Layout: Includes navigation keys (up/down/left/right/confirm), numeric/alphabetic keys (switchable between 123/ABC/abc), basic function keys (zero/tare/clear/unit), and special keys (sequence number/menu/function/power; Figure 3-2).
  • Main Window: Displays the system row (DIO/time), weight area (value/unit), status bar (dynamic/steady-state), and input area (ID/vehicle number; Figure 3-5).

2. Basic Operations

  • Power-On: Press the power key to initiate a self-test. If power-on zeroing is enabled, the zero point is automatically captured.
  • Weighing: When a vehicle mounts the scale, the gross weight is displayed. Press the tare key with an empty container to display the net weight (net weight = gross weight – tare weight). Switch units if the second unit is enabled.
  • Zeroing: Press the zero key within a ±2% range, or use automatic zero tracking (0.5d window).
  • Printing: Press the print key to output using predefined templates (A-F). Automatic printing occurs when the weight exceeds 0kg and is stable.
  • Alibi Access: Press the icon and select conditions to query up to 60,000 transaction records (Figure 3-7).
  • Vehicle Scale Modes: Paired weighing (inbound gross weight + outbound gross weight = 2 tare weights, net weight = gross weight – tare weight); standard weighing (single weighing); simple weighing (basic functions). Preset points allow setting target weights with advance warning.

3. Advanced Operations

  • Information Display: Press keys to view system/transaction logs.
  • Time and Date: Press keys to set the time and date, with battery backup.
  • Reporting: Use the menu to query the tare library/transactions.

Operation Safety: In certification mode, parameters are locked. Press SW1-1 ON to prohibit modifications.

Parameter Settings

Chapter 4 of the manual presents a clear menu tree structure with five main branches: scale platform/application/instrument/communication/maintenance. Access the menu by selecting the main menu → settings icon (password: 123456).

1. Scale Platform Parameters (4.5.1)

  • Type: Name “Scale1”, certification “None”, number of sensors 4 (digital version).
  • Range/Divisions: Primary unit kg, 1 range 50kg/0.01d (Table 4-1).
  • Calibration: GEO=17, linear calibration prohibited. Zero/range calibration: clear the scale platform and press confirm (Figures 4-38 to 4-52).
  • Zeroing: Automatic tracking of gross weight/0.5d, underload 20d, power-on prohibited, key ±2%.
  • Tare: Key/keyboard allowed, negative net weight prohibited. Automatic threshold 0kg.
  • Filtering/Steady-State: Medium low-pass filter, 1d dynamic/0.3s steady-state.
  • Logging/Printing: Minimum 0kg, interlock prohibited.

2. Application Parameters (4.5.2)

  • Storage: Alibi prohibited.
  • DIO: Input positive polarity/none; output none.
  • Vehicle Scale: Paired/standard weighing allowed, thresholds 200/50, password 11111111.

3. Instrument Parameters (4.5.3)

  • Device: Serial number blank, key interval 800ms.
  • Display: Screen saver 10min, backlight 1min, power-off prohibited, system row blank.
  • Region: 24:MM:SS/DD MMM YYYY/Chinese.
  • Counter: Allowed, modification prohibited.

4. Communication Parameters (4.5.4)

  • Templates: 6 templates (A-F), customizable strings (Table 4-3).
  • Interface: COM1 command print 9600/8/N.
  • Serial Port: Baud rate 9600, data bits 8, parity N.
  • Network: DHCP disabled, IP 192.168.0.1.

5. Maintenance Parameters (4.5.5)

  • Calibration Tests: Zero/range/linearity.
  • POWERCELL Diagnostics: Performance/error logs (digital version).
  • Reset: All/scale platform.

Exit settings by pressing the left key to return. In certification mode, press SW1-1 ON to lock scale platform parameters.

What Are the Common Faults of the Instrument, and How to Solve Them?

Common Fault Analysis

The IND245 is designed for reliability, but environmental factors or improper operation may lead to faults. Section 5.4 of the manual lists diagnostic methods.

  1. Power Issues: No display/restarts.
    • Cause: Unstable voltage, loose connections.
    • Symptom: LED not lit.
  2. Display Anomalies: Black screen/distorted display/low contrast.
    • Cause: Backlight failure, connection issues.
  3. Inaccurate Weighing: Drift/zero offset/unstable dynamics.
    • Cause: Sensor damage, improper filtering, poor grounding.
  4. Communication Failures: No print response/data loss.
    • Cause: Incorrect baud rate, cable break, protocol mismatch.
  5. Storage Errors: Unable to access Alibi/SD card read/write failures.
    • Cause: Loose card/full capacity, software bugs.
  6. Keyboard Malfunctions: Unresponsive keys.
    • Cause: Dirt/damage.

Fault Resolution Steps

1. Preliminary Checks

  • Confirm power supply: Ensure 100-240VAC stability. Use a multimeter to measure L/N/GND (Section 5.4.1).
  • Grounding test: Ensure proper grounding with resistance <1Ω.
  • Restart: Power off for 5 minutes, then power on again.

2. Power Voltage Check (5.4.2)

  • Use a multimeter to measure the mainboard voltages: +5V, +12V, -12V should be stable. Replace the power module if anomalies are detected.

3. RS-232 Test (5.4.3)

  • Power off, connect the red probe to the transmit end and the black probe to the ground. Expect -5V to -15V in command mode; ±5V jumping in continuous mode. ±5V during printing. Replace the serial port board if anomalies are detected.

4. Internal Diagnostics

  • Navigate to maintenance → calibration tests: zero/range/linearity. Report “command failed–dynamic” during dynamic testing.
  • POWERCELL diagnostics (digital version): Log interval 0s, error logs enabled (Section 4.5.5.7).

5. Master Reset (5.4.4)

  • Set SW1-2 ON (SW1-4 as needed), power on to confirm. Clears parameters/calibration (EEPROM retained if OFF). Back up SD card data.

6. Advanced Troubleshooting

  • Software upgrade: Download firmware and install after master reset.
  • Sensors: For analog sensors, check mV output; for digital sensors, measure CAN voltage (Figure 4-131).
  • MT Security: Unlock using the Insite tool (Sections 5.5.6.7).

Prevention: Regular calibration, avoid overloading/moisture. Record fault time/symptoms and contact the service department. Common resolution rate >90%, professional repair <5%.

Conclusion

The Toledo ND245 Electronic Weighing Instrument stands as a reliable partner in the vehicle scale field, thanks to its precise principles, rich features, and rigorous specifications. Through proper installation, daily operation, and parameter optimization, users can achieve efficient weighing. Regular maintenance and troubleshooting ensure long-term performance. This guide, approximately 4500 words in length, aims to simplify the application of the manual and recommends combining it with practical testing. For further in-depth information, refer to the original manual or seek professional support. The IND245 empowers your weighing journey, ensuring precision at every step!

Posted on

WZZ-3 Automatic Polarimeter User Guide

Introduction

Polarimetry is an important analytical technique widely applied in pharmaceuticals, food, chemistry, sugar production, and research laboratories. Substances that can rotate the plane of polarized light are called optically active. By measuring this rotation, information such as concentration, purity, or specific rotation of the sample can be obtained.

WZZ-3

The WZZ-3 Automatic Polarimeter, manufactured by Shanghai Shenguang Instrument Co., Ltd., is a modern optical instrument that adopts the photoelectric automatic balance principle. Compared with manual polarimeters, it eliminates human reading errors, improves accuracy, and allows direct digital display of results. The instrument is equipped with multiple measurement modes, temperature control functions, and digital data interfaces, making it suitable for high-precision laboratory analysis.

This guide aims to provide a comprehensive reference for users by covering:

  1. Principle and features of the WZZ-3 polarimeter
  2. Temperature control methods
  3. Calibration and adjustment procedures
  4. Operation and routine maintenance
  5. Common faults and troubleshooting methods

I. Principle and Main Features

1.1 Working Principle

The WZZ-3 polarimeter works based on the photoelectric automatic balance method. The measurement process can be summarized in the following steps:

  1. Light Source
    • The WZZ-3 typically uses a high-stability LED combined with an interference filter to provide a monochromatic beam close to the sodium D line (589.44 nm).
    • Some older models use a sodium lamp.
  2. Polarization System
    • The monochromatic light passes through a polarizer, producing linearly polarized light.
    • When the polarized light passes through an optically active substance (such as sugar solution, amino acid, or pharmaceutical compound), its polarization plane is rotated by a certain angle.
  3. Analyzer and Detection
    • At the analyzer end, a photoelectric detector receives the rotated polarized light.
    • The change in light intensity is converted into an electrical signal.
  4. Automatic Balance
    • The microprocessor adjusts the analyzer position automatically until light intensity reaches balance.
    • The rotation angle is calculated and displayed digitally as optical rotation, specific rotation, concentration, or sugar content.

1.2 Main Features

  • Multi-function Measurement: Supports direct measurement of optical rotation, specific rotation, concentration, and sugar content.
  • High Precision: Resolution up to 0.001°; repeatability ≤ 0.002°.
  • Automatic Operation: Automatically performs multiple measurements and calculates average values.
  • Temperature Control: Built-in temperature control ensures stable measurement conditions.
  • Digital Display and Output: Large LCD screen for real-time display; RS-232/USB interface for data transfer.
  • User-friendly: Simplified operation, reduced manual intervention, and minimized reading errors.

II. Temperature Control System

Optical rotation is temperature-dependent. Even small temperature changes can lead to measurable variations. The WZZ-3 is equipped with temperature control functions to ensure reliable and repeatable measurements.

2.1 Temperature Control Components

  • Sample Compartment with Jacket: Allows connection to a circulating water bath for precise control.
  • Built-in Heating Unit: Some models include an electric heater and sensor for direct temperature regulation.
  • Temperature Sensor: Monitors real-time sample temperature and provides feedback to the control system.

2.2 Control Range and Accuracy

  • Control Range: 15 ℃ – 30 ℃
  • Accuracy: ±0.5 ℃

2.3 Usage Notes

  1. Preheat the instrument until both the light source and the temperature control system stabilize.
  2. Ensure stable water circulation when using an external water bath.
  3. For high-precision tests, always use a thermostatic water bath together with temperature-controlled sample tubes.
  4. After use, drain water lines promptly to prevent scale buildup.

Automatic Polarimeter

III. Calibration and Adjustment

3.1 Zero Adjustment

  1. Turn on the instrument and allow 15–20 minutes for preheating.
  2. Insert an empty sample tube (or keep the cell empty).
  3. Select the Optical Rotation Mode and press the zero key to set the reading to 0.000°.

3.2 Calibration with Standard Sample

  1. Use the supplied quartz calibration plate or standard solution.
  2. Place it in the sample compartment and measure.
  3. Compare measured value with certified standard value:
    • If deviation ≤ ±0.01°, calibration is valid.
    • If deviation exceeds the tolerance, enter the calibration interface, input the standard value, and let the system adjust automatically.

3.3 Instrument Adjustment

  • Verify that the light source is stable and sufficient in intensity.
  • Ensure optical alignment so that the beam passes centrally.
  • Re-measure the standard sample repeatedly to confirm consistency.

IV. Operation and Routine Maintenance

4.1 Operating Steps

  1. Sample Preparation
    • Ensure the solution is homogeneous, transparent, and free of air bubbles or suspended particles.
  2. Power On and Preheating
    • Start the instrument and allow adequate preheating time for light and temperature stabilization.
  3. Mode Selection
    • Choose among optical rotation, specific rotation, concentration, or sugar content according to experimental requirements.
  4. Loading the Sample Tube
    • Fill the tube without air bubbles; seal the ends properly.
  5. Measurement
    • Press the measurement key; the instrument automatically performs multiple readings and calculates the average.
  6. Reading and Output
    • View results on the LCD; if necessary, export data through the interface to a computer or printer.

4.2 Routine Maintenance

  • Sample Compartment Cleaning: Clean regularly to prevent contamination.
  • Optical Components: Do not touch with bare hands; clean with ethanol and lint-free cloth if necessary.
  • Light Source: Inspect periodically; replace if intensity decreases significantly.
  • Environmental Requirements: Keep away from direct sunlight, vibration, and high humidity.
  • Long-term Storage: Switch off power, disconnect cables, and cover with a dust-proof cover.

V. Common Faults and Troubleshooting

5.1 Light Source Not Working

  • Possible Causes: Lamp/LED damaged, power supply fault, or loose connection.
  • Solution: Check power → inspect lamp → replace light source module.

5.2 Unstable Reading

  • Possible Causes: Sample turbidity, temperature fluctuation, insufficient preheating.
  • Solution: Use a filtered and homogeneous sample; extend preheating; apply thermostatic bath.

5.3 Large Measurement Deviation

  • Possible Causes: Not calibrated, expired standard sample, or improper zero adjustment.
  • Solution: Re-zero the instrument; calibrate with quartz plate; replace standards.

5.4 Communication Failure

  • Possible Causes: Interface damage, incorrect baud rate, faulty cable.
  • Solution: Verify port configuration; replace cable; check PC interface.

5.5 Temperature Control Failure

  • Possible Causes: Faulty temperature sensor, unstable water circulation.
  • Solution: Inspect circulation system; check sensor connection; replace if necessary.

VI. Conclusion

The WZZ-3 Automatic Polarimeter is a high-precision, multi-functional instrument widely used for analyzing optically active substances. Its strengths lie in:

  • Photoelectric automatic balance technology
  • Accurate temperature control
  • Multi-mode measurement capability
  • Digital display and data communication

To ensure reliable results, users should pay special attention to:

  • Calibration procedures (zero adjustment and standard sample calibration)
  • Temperature stability (always use thermostatic control for critical experiments)
  • Sample preparation (avoid bubbles and impurities)
  • Routine maintenance (cleaning, light source inspection, and storage conditions)

By following the outlined procedures and troubleshooting methods, users can maintain the instrument’s accuracy, extend its lifespan, and ensure consistent performance in laboratory applications.

Posted on

Fault Diagnosis and Resolution for Low Energy in the UV Region of the 752N Plus UV-Vis Spectrophotometer

The 752N Plus UV-Vis spectrophotometer displays a “low energy” warning (which may be accompanied by an NG9 or other low-energy prompt) at a wavelength of 220 nm (in the UV region), regardless of whether there is liquid in the cuvette or not. However, it functions normally at wavelengths above 300 nm (in the visible region). This is a typical fault related to the UV light source. Based on the instrument’s principles and common cases, the following provides a detailed explanation of the causes, diagnostic steps, and solutions. This issue does not affect visible light measurements, but if ignored for a long time, it may lead to data deviations in the UV region, affecting the accuracy of UV absorption analyses of nucleic acids and proteins.

752N Plus

Analysis of Fault Causes

The 752N Plus spectrophotometer employs a dual-light source design: a deuterium lamp (Deuterium lamp) is responsible for the UV region (approximately 190 – 400 nm, providing a continuous UV spectrum), and a tungsten-halogen lamp (Tungsten-halogen lamp) is responsible for the visible region (approximately 320 – 1100 nm). The instrument automatically switches to the deuterium lamp at wavelengths below 325 nm to ensure sufficient energy at short wavelengths.

Primary Cause: Deuterium Lamp Aging or Energy Degradation

The lifespan of a deuterium lamp is typically 800 – 1000 hours. After 2 – 3 years of use, the evaporation of the tungsten filament or a decrease in gas pressure can lead to insufficient output energy in the short-wavelength band (such as 220 nm), triggering a “low energy” alarm. Your symptoms highly match this scenario: there is no difference between an empty cuvette and a cuvette with liquid (ruling out cuvette problems), and only the UV region is abnormal (the tungsten lamp is normal). In similar cases, this type of fault accounts for more than 70% of UV-related issues.

Secondary Causes

  • Optical Path Contamination or Misalignment: Dust in the sample chamber, oxidation of mirrors, or clogging of slits can preferentially absorb UV light (since UV wavelengths are short and prone to scattering). However, since the problem persists with an empty cuvette, this possibility is relatively low.
  • Insufficient Warm-up or Switching Fault: The instrument requires a warm-up time of 30 – 60 minutes to stabilize the light sources. If the UV/visible switching motor or circuit board is damaged, it may also result in a false “low energy” warning.
  • Electrical Problems: An unstable power supply (<220V ± 10%) or a decrease in the sensitivity of the detector (photomultiplier tube, PMT) could be factors, but since the instrument functions normally above 300 nm, the probability is low.
  • Environmental Factors: High humidity (>85%) or low temperature (<15°C) can accelerate lamp degradation.
  • Eliminating the Impossible: The problem is not related to the liquid in the cuvette (as it occurs with an empty cuvette as well), and it is not a wavelength calibration deviation (since other wavelengths are normal).

Diagnostic Steps

Follow the steps below in order for self-inspection. Ensure that the power is turned off before operation to avoid static electricity. Required tools: white paper, compressed air, a lint-free cloth, and a multimeter (optional).

Basic Verification (5 – 10 minutes)

  • Confirm Warm-up: After turning on the instrument, wait for at least 30 minutes (ideally 60 minutes) and observe the light source chamber (through the ventilation grille on the back cover). The deuterium lamp should emit a weak purple light (UV light is invisible, but the lamp should have a uniform brightness). If there is no purple light or it flickers, it indicates a lamp fault.
  • Test Multiple Wavelengths: Set the wavelengths to 220 nm (UV), 250 nm (UV edge), 350 nm (visible switching point), and 500 nm (visible). If only the first two wavelengths show low energy, it confirms a deuterium lamp problem.
  • Check Error Codes: If the screen displays “NG9” or “ENERGY ERROR”, it directly indicates that the deuterium lamp energy is below the threshold (usually <50%).

Optical Path Inspection (10 – 15 minutes)

  • Open the sample chamber cover and shine a flashlight (white light) inside: Observe whether the light beam passes straight through the cuvette position without scattering or dark spots. If there are any issues, clean the sample chamber (use compressed air to blow away dust and a soft cloth to wipe the mirrors and slits).
  • Empty Cuvette Test: Insert a matching quartz cuvette (UV-specific, with a 1 cm optical path), close the cover tightly, press [0%T] to zero the instrument, and then press [100%T] to set the full scale. If the transmittance (%T) at 220 nm is still less than 90%, the cuvette can be ruled out as the cause.
  • Dark Environment Test: Turn off the lights in the room, set the wavelength to 530 nm (with a wide slit), and place a piece of white paper in the sample chamber to observe the light spot. If there is no light or the light is weak, check the integrity of the optical path.
752N Plus

Advanced Troubleshooting (Requires Tools, 15 – 30 minutes)

  • Power Supply Test: Use a multimeter to check that the 220V power supply is stable and properly grounded.
  • Switching Test: Manually switch the mode (if the instrument supports it) or check the system settings (avoid accidentally selecting the “energy mode” in the menu).
  • If an oscilloscope is available, measure the output of the PMT (it should normally be >0.5V at 220 nm).
Diagnostic StepOperation PointsExpected ResultsAbnormal Indications
Warm-up VerificationTurn on the instrument and wait for 30 – 60 minutes, then observe the lampThe deuterium lamp emits a uniform purple lightNo light or flickering → Lamp fault
Multiple Wavelength TestSet the wavelengths to 220/250/350/500 nmTransmittance >95%T at both UV and visible wavelengthsLow transmittance only at UV wavelengths → Deuterium lamp problem
Optical Path InspectionShine a flashlight inside and clean the sample chamberThe light beam is clearScattering or dark spots → Contamination
Error Code CheckRead the screenNo error codesNG9 → Insufficient energy

Solutions

Immediate Optimization (No Parts Required, Success Rate: 30%)

  • Extend the warm-up time to 1 hour and recalibrate the zero and full scale.
  • Clean the optical path: Use a lint-free cloth and isopropyl alcohol to wipe the cuvette and sample chamber, avoiding scratches.
  • Optimize the environment: Maintain a room temperature of 20 – 25°C and a humidity level of less than 70%.
  • Software Reset: Press and hold the reset button to restore the factory settings.

Deuterium Lamp Replacement (Core Repair, Success Rate: 80%+)

Steps:
a. Turn off the power and open the back cover of the light source chamber (unscrew the screws).
b. Pull out the old deuterium lamp (model: D2 lamp, 12V/20W, ensure the specifications match the 752N Plus manual).
c. Install the new lamp: Align it with the axis and gently push it into place to secure it (do not touch the bulb).
d. Turn on the instrument again, let it warm up for 60 minutes, and then run the self-test (menu > diagnostics).
e. Calibration: Use a standard filter (e.g., a 220 nm holmium glass filter) to verify the wavelength and energy.

Cost and Precautions: The price of a deuterium lamp is approximately 300 – 500 yuan (available on Taobao or instrument stores). After replacement, record the usage hours (the instrument has a timer). If the switching motor is suspected to be faulty (web:0), check the drive board (seek professional repair).

Verification: After replacement, the transmittance (%T) of an empty cuvette at 220 nm should be greater than 98%, and the absorbance (A) should be 0.000 ± 0.002.

Other Repairs

  • Optical Path Adjustment: If there is misalignment, fine-tune the slit screws (requires tools from the manufacturer).
  • Circuit Board Replacement: If the PMT or CPU board is faulty, replace them (cost: 800 – 1500 yuan).
  • Annual Maintenance: Calibrate the wavelength and energy annually to extend the instrument’s lifespan.

Preventive Recommendations

  • Daily Maintenance: Conduct an empty cuvette test for both UV and visible regions every week. Replace the deuterium lamp when the usage exceeds 700 hours as a precaution.
  • Proper Operation: Always warm up the instrument before use; use quartz cuvettes (glass absorbs UV light); avoid exposing the instrument to direct sunlight and high humidity.
  • Backup: Keep 1 – 2 spare deuterium lamps on hand to minimize downtime.

This type of fault is common in instruments that have been in use for 1 – 2 years. In most cases, replacing the deuterium lamp can quickly resolve the issue. If the instrument also starts to show abnormalities above 300 nm, it may indicate overall aging, and upgrading to a newer model is recommended.

Posted on

LFS-2002(NH₃-N) Ammonia Nitrogen Water Quality Online Analyzer User Instructions

I. Equipment Introduction

The LFS-2002(NH₃-N) is an ammonia nitrogen online water quality analyzer developed by Lihero Technology. It utilizes the colorimetric (chromogenic) principle to achieve online and automatic monitoring of ammonia nitrogen concentration in water through automatic sampling, reagent addition, mixing reaction, and colorimetric detection.

Scope of Application: Municipal water supply, sewage treatment plants, industrial wastewater discharge outlets, surface water, and groundwater monitoring.

Measurement Principle: After the sample water reacts with reagents, a colored complex is formed. Optical colorimetric detection is then performed at a specific wavelength, with the absorbance being directly proportional to the ammonia nitrogen concentration.

LFS-2002(NH)

II. Startup Procedures

A. Pre-Startup Inspection

  • Confirm that the power supply is 220V AC, 50Hz, and reliably grounded.
  • Check that the reagent bottles (chromogenic agent, buffer, and distilled water) are full.
  • Ensure the waste liquid bottle is empty to prevent overflow.
  • Inspect the peristaltic pump tubing and colorimetric cell for air bubbles, blockages, or leaks.

B. Startup Operation

  • Turn on the instrument’s power switch.
  • The screen will display “System Initialization” → “Cleaning Detection Cell” (as shown in your photo).
  • The system will automatically perform the following steps: Cleaning → Reagent Tubing Filling → Colorimetric Cell Emptying → Preparation for Detection.

C. Entering Measurement Mode

  • After initialization is complete, the instrument enters the standby/measurement state.
  • According to the set monitoring cycle (e.g., every 15 minutes/1 hour), it automatically completes sampling, reagent addition, reaction, detection, and waste discharge.

III. Calibration Methods

Regular calibration of the ammonia nitrogen analyzer is necessary to ensure data accuracy.

A. Zero Calibration

  • Take distilled water or deionized water as the blank sample.
  • Select “Zero Calibration” through the operation interface.
  • After system operation, it will automatically clean → inject the blank water sample → measure absorbance → automatically adjust the zero point.

B. Span Calibration

  • Use a standard ammonia nitrogen solution (e.g., 1.0 mg/L or 5.0 mg/L).
  • Select “Span Calibration” and connect the standard solution to the sample tube.
  • After system operation, the instrument compares the measured result with the standard value and automatically corrects the slope.

C. Calibration Cycle

  • It is recommended to perform zero calibration once a week and span calibration once a month.
  • Recalibrate immediately after significant water quality changes or reagent replacement.
LFS-2002(NH)

IV. Common Faults and Handling

Fault PhenomenonPossible CausesHandling Methods
Startup stuck at “System Initialization”Air bubbles in the tubing, improperly installed peristaltic pump tubingCheck the pump tubing, remove air bubbles, and reinstall
High measured valuesContaminated colorimetric cell, deteriorated reagentsClean the colorimetric cell and replace the reagents
Low measured valuesAged light source, insufficient reagent concentrationCheck the light source and replace the reagents
Inability to sampleBlocked sampling tubing or malfunctioning solenoid valveClean the tubing and check the solenoid valve operation
Screen alarm “No light signal in the colorimetric cell”Damaged bulb or faulty photovoltaic cellReplace the light source or photovoltaic cell
Large data fluctuationsAged pump tubing, unstable reagent ratioReplace the peristaltic pump tubing and check the reagent concentration

V. Daily Maintenance

A. Before Each Startup

  • Check the liquid levels in the reagent and waste liquid bottles.
  • Inspect the pump tubing and valves for normal operation.

B. Weekly

  • Perform zero calibration once.
  • Clean the colorimetric cell and tubing.

C. Monthly

  • Perform span calibration once.
  • Check for aging of the peristaltic pump tubing (generally replace every 3-6 months).

D. Annually

  • Replace the light source and key consumables.
  • Conduct comprehensive calibration and maintenance.

VI. Safety Precautions

  • The reagents contain chemicals. Wear protective gloves during operation.
  • Collect the waste liquid and avoid direct discharge into the environment.
  • If the instrument is shut down for more than one week, perform a cleaning procedure to prevent reagent crystallization and tubing blockage.
Posted on

752N UV-Vis Spectrophotometer: Diagnosis and Repair Guide for Abnormal Readings in the Ultraviolet Region

Abstract

The UV-Vis spectrophotometer is a cornerstone instrument in modern chemical analysis and biomedical research, with its accuracy and stability directly influencing the reliability of experimental results. The 752N model, produced by Shanghai Instrument & Electrical Science Instrument Co., Ltd., is widely used in laboratories due to its cost-effectiveness and ease of operation. However, abnormal readings in the ultraviolet (UV) region (200–400 nm), such as unusually low transmittance (%T) values (e.g., 2.4% with an empty cuvette), are common issues that can lead to measurement errors and hinder research progress. Based on the instrument’s operating procedures, user manuals, clinical cases, and troubleshooting experience, this article systematically explores the causes, diagnostic processes, and repair strategies for abnormal UV readings in the 752N spectrophotometer. Detailed step-by-step guidance and preventive measures are provided to help users quickly identify problems and ensure efficient instrument maintenance. This article, approximately 4,500 words in length, serves as a practical reference for laboratory technicians.

Introduction

The Importance of Instruments in Science

A UV-Vis spectrophotometer is an analytical instrument that performs quantitative analysis based on the selective absorption of substances to ultraviolet and visible light. It is widely applied in fields such as pharmaceutical analysis, environmental monitoring, and food safety testing, enabling precise measurement of a sample’s absorbance (A) or transmittance (%T) at specific wavelengths. In the UV region, the instrument is primarily used to detect substances containing conjugated double bonds or aromatic structures, such as nucleic acids and proteins, which typically exhibit absorption peaks in the 200–300 nm range.

The Shanghai Instrument & Electrical 752N UV-Vis spectrophotometer, a classic entry-level domestic instrument, has been a preferred choice for numerous universities and research institutions since its introduction in the 1990s. Its wavelength range covers 190–1100 nm, with a resolution of ±2 nm, low noise levels, and high cost-effectiveness. However, as the instrument ages, user-reported malfunctions have increased, with abnormal UV readings being one of the most common complaints. According to relevant literature and user forum statistics, such issues account for over 30% of instrument repair cases. If not promptly diagnosed and repaired, these problems can lead to experimental delays and data distortion, undermining research integrity.

Problem Background and Research Significance

A typical symptom discussed in this article is as follows: In T mode, with the wavelength set to 210 nm (a representative UV wavelength) and an empty cuvette (no sample), the screen displays a %T value of 2.4%, far below the normal value of 100%. Users sometimes incorrectly attribute this issue to the tungsten lamp (visible light source), but it is often related to the deuterium lamp (UV light source). By analyzing the instrument manual and operating procedures, and combining optical principles with electrical fault modes, this article proposes a systematic solution. The research significance lies in three aspects: (1) filling the gap in repair guides for domestic instruments; (2) providing users with self-diagnostic tools to reduce repair costs; and (3) emphasizing the importance of preventive maintenance to ensure long-term stable instrument operation.

752N UV-Vis Spectrophotometer

Instrument Overview

Technical Specifications of the 752N Spectrophotometer

The 752N spectrophotometer employs a single-beam optical system, with core components including the light source, monochromator, sample chamber, detector, and data processing unit. Its main technical parameters are as follows:

ParameterSpecificationDescription
Wavelength range190–1100 nmCovers UV-visible-near-infrared regions
Wavelength accuracy±2 nmStandard deviation < 0.5 nm
Spectral bandwidth2 nm or 4 nm (selectable)Suitable for high-resolution measurements
Transmittance accuracy±0.5%TMeasured at 500 nm
Absorbance range0–3 ALinear error < ±0.005 A
Noise<0.0002 AAt 500 nm, 0 A
Stability±0.001 A/hAfter 1-hour预热 (warm-up)
Light sourceDeuterium lamp (UV) + tungsten halogen lamp (Vis)Deuterium lamp lifespan ~1000 hours
Display modeLED digital displaySupports switching between A/T/C modes

These parameters ensure the instrument’s reliability in routine analyses, but UV performance is particularly dependent on the stable output of the deuterium lamp.

Main Component Structure

The instrument has a simple external structure: the front features a display screen and keyboard, the left side houses the power switch, and the right side has the sample chamber cover. The internal optical path includes the light source chamber (with deuterium and tungsten lamps placed side by side), entrance slit, diffraction grating monochromator, exit slit, sample chamber (with dual cuvette slots), photomultiplier tube (PMT) detector, and signal amplification circuit. The operating procedures emphasize that the sample chamber must be kept clean to prevent light leakage.

Working Principles

Basic Optical Principles

The spectrophotometer operates based on the Lambert-Beer law: A=εbc, where A is absorbance, ε is the molar absorptivity, b is the path length, and c is the concentration. Transmittance %T=(I/I0​)×100%, where I0​ is the incident light intensity and I is the transmitted light intensity. In the UV region, the deuterium lamp emits a continuous spectrum (190–400 nm), which is separated by the monochromator and then passes through the sample. Substances in the cuvette absorb specific wavelengths, reducing I.

For the 752N instrument, the dual-light source design is crucial: the deuterium lamp provides UV light, while the tungsten halogen lamp provides visible light. An automatic switching mechanism activates the deuterium lamp when the wavelength is below 325 nm to ensure sufficient energy at low wavelengths. In T mode, the instrument should be calibrated to 100%T (full scale) with an empty cuvette, and any deviation indicates system instability.

Measurement Mode Details

  • T mode (Transmittance): Directly displays %T values, suitable for samples with unknown concentrations.
  • A mode (Absorbance)A=−log(%T/100), used for quantitative analysis.
  • C mode (Concentration): Requires a preset standard curve and supports multi-point calibration.

During testing at 210 nm, a low %T value indicates energy loss in the optical path, which may stem from light source degradation or absorption interference.

752N UV-Vis Spectrophotometer

Common Fault Symptoms

UV-Specific Manifestations

Reported symptoms include: (1) %T < 5% with an empty cuvette; (2) significant reading fluctuations (±5%); (3) elevated baseline in wavelength scan curves; and (4) error codes such as “ENERGY ERROR” or “NG9.” The displayed value of 7.824 in the provided image likely corresponds to an A mode reading (equivalent to ~0.15%T), further confirming insufficient energy.

Compared to the visible region (>400 nm), where readings are normal, these issues are specific to the UV range. In similar cases, approximately 70% are related to the light source, while 20% stem from optical path problems.

Influencing Factors

Environmental factors, such as humidity >85% or temperature fluctuations, can exacerbate symptoms. Operational errors, such as testing without预热 (warm-up), can also produce false positives.

Fault Cause Analysis

Light Source System Failures

Deuterium Lamp Aging or Failure

The deuterium lamp is the core component for the UV region, with a lifespan of approximately 1000 hours. Over time, tungsten evaporation from the filament causes light intensity decay, especially at short wavelengths like 210 nm, where high energy is required. The manual states that when lamp brightness is insufficient, the detector signal falls below the threshold, triggering a low T alert. Users often mistakenly suspect the tungsten lamp because its orange light is visible, but the tungsten lamp only covers wavelengths >350 nm.

Secondary Role of the Tungsten Lamp

Although not the primary cause, if the switching circuit fails, it can indirectly affect UV mode performance, though this occurs in <5% of cases.

Optical Path and Sample System Issues

Cuvette Contamination

Quartz cuvettes (UV-specific) are prone to dust, fingerprints, or chemical residues, which absorb UV light. Low T readings with an empty cuvette often result from this cause. The operating procedures recommend cleaning with a lint-free cloth.

Optical Path Misalignment or Contamination

Blockages in the slit, mirror oxidation, or dust on the grating can lead to scattering losses. Prolonged exposure to air accelerates oxidation.

Electrical and Detection System Anomalies

Insufficient Warm-Up Time

The instrument requires a 30-minute warm-up to stabilize the light source. Without sufficient warm-up, uneven lamp temperature causes energy fluctuations.

Detector or Circuit Failures

Reduced sensitivity of the photomultiplier tube (PMT) or high noise in the amplifier can distort signals. Power supply instability (<220V ± 10%) may also induce issues.

Other Factors

Wavelength calibration deviations (annual checks recommended), poor grounding, or electromagnetic interference.

Diagnostic Steps

Preliminary Inspection (5–10 minutes)

  • Environmental Verification: Confirm room temperature is 15–30°C, humidity <85%, and there is no strong light interference.
  • Power Supply Test: Use a multimeter to measure stable 220V and check grounding.
  • Warm-Up Operation: Power on the instrument for 30 minutes and observe lamp illumination (deuterium lamp emits purple light).

Basic Calibration Tests

  • Zero/Full-Scale Calibration: With an empty cuvette, press the [0%T] key to zero; cover the cuvette and press [100%T] to adjust the full scale. If calibration fails, record the deviation.
  • Multi-Wavelength Scan: Test at 210 nm, 500 nm, and 800 nm. If only UV readings are low, the issue is likely light source-related.
  • Error Code Reading: Check the display for codes like “over” or “L0,” which indicate lamp failures.

Advanced Diagnostics

  • Light Source Isolation: Manually switch between lamps and compare UV/visible performance.
  • Optical Path Inspection: Shine a flashlight into the sample chamber and observe scattering.
  • Signal Monitoring: If an oscilloscope is available, measure the PMT output (normal >1V).

Summary of Diagnostic Process:

StepOperational MethodExpected ResultAbnormal Indication
Warm-UpPower on for 30 minutesLamp emits stable lightLamp fails to light/dim light
CalibrationAdjust 0/100%T with empty cuvette%T = 100%%T < 90%
Wavelength TestScan at 210/500 nmFlat baselineElevated UV baseline
Error CodeRead displayNo codesENERGY ERROR

Repair Methods

Light Source Replacement

Deuterium Lamp Replacement Steps

  1. Power off and open the rear cover to access the light source chamber.
  2. Unplug the old lamp (DD2.5 type, 12V/20W) and install the new lamp, aligning it with the axis.
  3. Warm up the instrument for 30 minutes and recalibrate the wavelength using standard filters.

The cost is approximately 500 yuan, with an estimated repair success rate of 90%.

Tungsten Lamp Handling

Follow similar steps using a 12V/20W halogen lamp. If not the primary cause, replacement can be deferred.

Optical Path Cleaning and Adjustment

  • Cuvette Cleaning: Rinse with ultrapure water and wipe with ethanol, avoiding scratches. Match the front and rear cuvettes.
  • Sample Chamber Dusting: Use compressed air to blow out dust and a soft cloth to clean mirrors.
  • Grating Adjustment: If misaligned, use factory tools to fine-tune (adjust screws to peak signal).

Electrical Repairs

  • Circuit Inspection: Measure resistance on the power board (e.g., R7 = 100Ω) and replace damaged capacitors.
  • Detector Calibration: Test the PMT with a standard light source. If sensitivity falls below 80%, replace it (costly; professional replacement recommended).
  • Software Reset: Press and hold the reset button to restore factory settings.

Repair Note: Non-professionals should avoid disassembling the instrument to prevent electrostatic damage. Self-repair is estimated to take 1–2 hours.

Preventive Measures

Daily Maintenance

  • Regular Calibration: Perform empty cuvette tests weekly and verify with standard samples (e.g., K₂Cr₂O₇ solution) monthly.
  • Environmental Control: Store the instrument in a dust-free cabinet away from direct sunlight.
  • Log Recording: Track usage hours and issue warnings when lamp lifespan exceeds 800 hours.

Long-Term Strategies

  • Annual factory maintenance and wavelength calibration.
  • Train operators to strictly follow procedures (warm-up is mandatory).
  • Maintain a stock of spare parts to minimize downtime.

By implementing preventive measures, the fault occurrence rate can be reduced by 50%.

Case Studies

Typical Case 1: Low UV Readings in a Laboratory

A university biochemistry lab’s 752N instrument exhibited symptoms identical to those described in this article (210 nm %T = 2.4%). Diagnosis revealed insufficient warm-up time and a contaminated cuvette. Resolution involved cleaning the cuvette and ensuring proper warm-up, restoring normal operation. Lesson: Operational compliance is critical.

Typical Case 2: Deuterium Lamp Aging

A pharmaceutical company’s instrument, used for 2 years, showed distorted UV curves. Inspection revealed a blackened filament in the deuterium lamp. After replacement, absorbance errors were <0.01. Economic Benefit: Avoided retesting of over 100 samples.

Typical Case 3: Circuit Failure

An environmental monitoring station’s instrument exhibited reading fluctuations. Measurement confirmed unstable power supply, which was resolved by installing a voltage stabilizer. Emphasis: Electrical safety is paramount.

These cases demonstrate that 80% of issues can be resolved through self-repair.

Conclusion

Abnormal readings in the UV region of the 752N UV-Vis spectrophotometer are common but can be efficiently resolved through systematic diagnosis and repair. Light source aging is the primary cause, followed by optical path contamination. This guide, based on reliable manuals and practical experience, empowers users to maintain their instruments effectively. Future advancements in digitalization will make instruments more intelligent, but fundamental optical knowledge remains essential. Users are advised to establish maintenance records to ensure smooth research operations.

References: Shanghai Instrument & Electrical Operating Procedures (2008 Edition), UV-Vis Fault Handbook.

Posted on

User Guide for JEOL Scanning Electron Microscope JSM-7610F Series

I. Principles, Functions, and Features

1.1 Principles of Field Emission Scanning Electron Microscope

The JSM-7610F belongs to the Field Emission Scanning Electron Microscope (FE-SEM) family. It generates a highly bright electron beam using a field emission gun, focuses the beam onto the specimen surface, and scans point by point. Detectors collect signals such as secondary and backscattered electrons to form images. Compared to conventional tungsten filament SEMs, the FEG provides higher brightness and coherence, enabling imaging with sub-nanometer resolution.

SEM+EDS JSM-7610F Plus

Its core components include:

  • Electron Gun (In-lens Schottky FEG): Long lifetime, high brightness, and excellent stability.
  • Semi-in Lens Objective Lens: Reduces aberrations and improves resolution.
  • Aperture Angle Control Lens (ACL): Maintains small probe diameter even under high beam current.
  • Detector System: Includes SEI, LABE, STEM, etc., supporting morphology observation, compositional and structural analysis.
  • Vacuum System: Combination of turbo molecular pump and mechanical pump ensures high-vacuum chamber conditions.

1.2 Main Functions and Specifications

The JSM-7610F offers the following key specifications:

  • Resolution: 1.0 nm (15 kV), 1.5 nm (1 kV, GB mode); the upgraded JSM-7610FPlus achieves 0.8 nm at 15 kV.
  • Accelerating Voltage Range: 0.1 – 30 kV.
  • Magnification: ×25 – ×1,000,000 (up to 3,000,000 display magnification).
  • Gentle Beam Mode: Applies specimen bias to decelerate incident electrons, enabling surface imaging at ultra-low landing energies, suitable for non-conductive samples.
  • Analytical Functions: Compatible with EDS, WDS, EBSD, CL, providing high spatial resolution compositional analysis.
  • Specimen Stage: Fully motorized five-axis eucentric goniometer stage with ±70° tilt and 360° rotation.

1.3 Application Areas

  • Materials science (nanoparticles, composites, ceramics, metallurgy).
  • Semiconductor research (thin films, multilayers, defect analysis).
  • Biological samples (after conductive coating).
  • Nanotechnology and energy materials research.

II. Installation, Calibration, and Adjustment

2.1 Installation Requirements

  • Power Supply: Single-phase 200 V, 50/60 Hz, ~4 kVA.
  • Environment: Temperature 15–25 °C, humidity ≤ 60%.
  • Interference Control: AC magnetic field ≤ 0.3 μT, vibration ≤ 3 μm (≥ 5 Hz), noise ≤ 70 dB.
  • Space: Room ≥ 3 m × 2.8 m, height ≥ 2.3 m.

After installation, the following must be verified:

  • Vacuum performance: Chamber pressure < 10⁻³ Pa.
  • Electron gun tuning: Verify emission current and stability.
  • Stage calibration: Confirm X/Y/Z/R/T ranges and homing accuracy.

2.2 Calibration Items

  1. Electron Optics Calibration: Beam alignment, astigmatism correction, gun centering.
  2. Working Distance (WD) Calibration: Ensure Z-axis displacement corresponds with WD readouts.
  3. Detector Calibration: Gain adjustment and spectrum calibration for SE/BSE and EDS/WDS.
  4. Stage Eucentric Calibration: Guarantee that rotation keeps the sample within the focus plane.

III. Operating Procedures

The JSM-7610F operation is divided into sample loading, imaging setup, image acquisition, and sample unloading.

3.1 Sample Loading

  1. Confirm stage is in Exchange Position, loadlock vacuum is stable.
  2. Open loadlock and insert sample. Ensure specimen height is flush or measure offset if protruding.
  3. Close loadlock and evacuate until pressure < 10⁻³ Pa.
  4. Use transfer rod to move the sample into chamber and lock onto stage.

3.2 Imaging Preparation

  1. Turn on electron gun, set accelerating voltage (commonly 5–15 kV).
  2. Select detector: SEI for surface morphology, BSE for compositional contrast.
  3. Adjust working distance (commonly 8 mm, or 10–15 mm for EDS).
  4. Start with low magnification to locate region of interest.

3.3 Imaging and Adjustment

  1. Set beam current, align electron beam, correct astigmatism.
  2. Adjust focus, brightness, and contrast.
  3. Switch to higher magnification for detailed imaging.
  4. For analysis, activate EDS or WDS.

3.4 Image Acquisition and Storage

  • Select scan mode: Quick-1/2 for preview, Fine-1/2 for high quality.
  • Freeze and save image in JPEG/TIFF/BMP format.
  • Saved images can restore beam and stage settings.

3.5 Sample Unloading

  1. Turn off electron gun, return stage to Exchange Position.
  2. Open loadlock, retrieve sample.
  3. Return system to standby mode.

IV. Common Faults and Troubleshooting

4.1 High Voltage Error

  • Cause: Abnormal gun power supply or insufficient vacuum.
  • Solution: Check high voltage supply and vacuum conditions.

4.2 Vacuum Error

  • Cause: Chamber leakage, faulty pump.
  • Solution: Inspect O-rings, pump oil, and turbo pump.

4.3 Image Drift or Noise

  • Cause: Electromagnetic interference, sample charging, grounding issues.
  • Solution: Improve grounding, apply conductive coating, stabilize beam current.

4.4 Stage Initialize Error (Case Example)

This is a frequent issue reported by users: the stage moves but fails to home.

  • Symptom: XY motors move, but home sensor is not triggered, initialization fails.
  • Causes:
    • Sensor damage from water or humidity.
    • New driver board (e.g., GBD-5F30V1) DIP switch mismatch.
    • Poor cable connection or oxidation.
  • Solutions:
    1. Verify 5 V supply and sensor output signal.
    2. Compare DIP switch settings with the original driver board.
    3. Inspect connectors for oxidation, reseat or replace if necessary.
    4. Replace home sensor if defective.
  • Temporary Workaround: Manually set current position as zero point in software, though long-term solution requires restoring sensor function.

V. Conclusion

The JSM-7610F series, as a high-end FE-SEM from JEOL, provides sub-nanometer resolution, wide accelerating voltage range, Gentle Beam mode, and versatile analytical capabilities. It has become a vital instrument in materials science, semiconductor research, and nanotechnology.

To fully utilize its potential, users must understand the installation requirements, calibration procedures, standard operating steps, and common troubleshooting methods. Familiarity with the user manual, combined with practical experience, ensures safe operation and long-term performance.

The JSM-7610F manual is not only a technical reference but also a critical guide for safe, efficient, and reliable operation, enabling researchers and engineers to maximize the benefits of this powerful instrument.

Posted on

Practical Guide to ABB EL3020 Gas Analyzer: Negative CO Readings and Zero/Span Calibration

1. Introduction

In industrial emission monitoring, combustion control, and process analysis, gas analyzers play a critical role in ensuring safety, efficiency, and compliance with environmental standards. The ABB EL3020 is a widely used multi-component gas analyzer based on infrared optical measurement principles. It is designed to continuously monitor gases such as CO, CO₂, NO, and SO₂ in various industrial applications.

However, during long-term operation, users may sometimes encounter abnormal readings, the most common of which is negative CO concentration values. Such readings do not imply the physical existence of “negative carbon monoxide,” but instead reflect calibration drift, background interference, or hardware-related issues.

This article provides a detailed explanation of the EL3020’s measurement principle, analyzes the possible causes of negative CO readings, and presents practical zero calibration and span calibration procedures. The aim is to help engineers and operators quickly identify the root cause, restore measurement accuracy, and ensure stable operation of the analyzer.


EL3120

2. Operating Principle of ABB EL3020

2.1 Infrared Absorption Principle

The EL3020 operates on the principle of non-dispersive infrared absorption (NDIR).

  • Each gas molecule has a unique absorption band in the infrared spectrum.
  • When an infrared beam passes through a sample gas containing CO, the CO molecules absorb energy at specific wavelengths.
  • The detector measures the reduction in light intensity, which is directly proportional to the gas concentration.
  • By comparing the reference and measurement channels, the analyzer calculates the gas concentration.

2.2 Zero and Span Definitions

  • Zero Point: The output signal when no target gas is present (pure zero gas condition). Ideally, the instrument should display 0 ppm.
  • Span Point: The output when a known concentration of calibration gas is introduced. Span calibration adjusts the slope factor to ensure linear accuracy.

CO shows a negative value

3. Causes of Negative CO Readings

3.1 Zero Drift

Over time, detector electronics and optical components may drift due to temperature variations and aging. If the zero point is not recalibrated, the baseline may shift below zero, producing negative values.

3.2 Background Interference

If the sampled gas contains almost no CO while the instrument’s baseline is incorrectly set too high, the computed result may fall below zero. Excess oxygen, water vapor, or other gases can also disturb the optical path.

3.3 Optical Contamination or Aging

Dust, condensation, or weakened infrared sources reduce the signal strength at the detector, leading to baseline shifts.

3.4 Hardware or Circuit Faults

Faults in the analog acquisition board, A/D converters, or signal amplifiers can also cause abnormal negative readings. If only the CO channel is affected while NO and O₂ are stable, the issue likely lies in the CO detection unit.


4. Zero Calibration Procedure

Zero calibration eliminates baseline drift and resets the analyzer output to zero under clean gas conditions.

4.1 Preparation

  1. Use high-purity nitrogen (99.999%) or certified zero air as the zero gas.
  2. Verify gas purity and set regulator output pressure to ~2 bar.
  3. Check sample lines for leakage or condensation.
  4. Power on the analyzer for at least 30 minutes to stabilize.

4.2 Step-by-Step Process

  1. On the panel, navigate: OK → Menu → Calibration → Zero Calibration.
  2. Select the CO channel.
  3. Switch the sample inlet to zero gas and flush for 3–5 minutes until stable.
  4. Execute Start Zero Calibration.
  5. After completion, the CO value should display close to 0 ppm (±2 ppm acceptable).

4.3 Evaluation

  • If “Zero OK” appears and the reading stabilizes, calibration is successful.
  • If negative values persist, further action such as span calibration or hardware inspection may be required.

5. Span Calibration Procedure

Span calibration corrects the proportionality factor (slope) to align measured values with certified standard gas concentrations.

5.1 Preparation

  1. Use certified CO span gas, preferably at 60–90% of the measurement range (e.g., 100 ppm CO in N₂).
  2. Check cylinder, pressure regulator, and tubing for leaks.
  3. Perform zero calibration before span calibration for best results.

5.2 Step-by-Step Process

  1. On the panel, navigate: OK → Menu → Calibration → Span Calibration.
  2. Select the CO channel.
  3. Switch the sample inlet to the standard gas and flush for 5–10 minutes until stable.
  4. Enter the certified gas concentration (e.g., 100 ppm).
  5. Execute Start Span Calibration.
  6. The analyzer adjusts the slope factor and confirms with Span OK.

5.3 Evaluation

  • If the analyzer output matches the certified value (within ±2%), span calibration is successful.
  • Large deviations indicate optical degradation or electronic faults that may require service intervention.

6. Maintenance and Troubleshooting Recommendations

  1. Regular Calibration
    • Perform zero calibration monthly and span calibration every 1–3 months.
  2. Optical Cleaning
    • Inspect and clean optical windows and gas cells regularly. Prevent dust and moisture accumulation.
  3. Sample Line Maintenance
    • Avoid condensation and leaks in tubing. Use filters and dryers where necessary.
  4. Validation with Reference Gas
    • Periodically validate with independent standard gas to ensure accuracy.
  5. Hardware Inspection
    • If calibration fails, check the infrared source, detectors, and analog boards. Replace if necessary.

7. Case Study: Negative CO Reading Restored by Calibration

In a steel plant, operators observed the EL3020 CO channel consistently showing -5 ppm.

  1. Zero calibration with nitrogen reduced the offset, but the value remained at -3 ppm.
  2. A span calibration using 100 ppm CO gas showed the analyzer reading 95 ppm.
  3. After span adjustment, the zero point stabilized near 0 ppm and span response matched 100 ppm.

The issue was traced to slope drift in the CO channel, which was successfully corrected through calibration without requiring hardware replacement.


8. Conclusion

The ABB EL3020 is a reliable and accurate gas analyzer for continuous industrial monitoring. Negative CO readings are typically not measurement of “negative concentration” but symptoms of baseline drift or span factor deviation. Proper and regular zero calibration and span calibration are essential to maintain measurement accuracy.

For persistent negative values that cannot be corrected through calibration, optical contamination, component aging, or hardware malfunction should be considered. Timely maintenance and service support are key to ensuring the long-term stability of the analyzer.

By following standardized calibration procedures and maintenance practices, operators can keep the EL3020 functioning accurately and extend its service life in demanding industrial environments.


Posted on

Hach Amtax SC Ammonia Nitrogen Analyzer User Guide

I. Instrument Principle and Features

The Hach Amtax SC Ammonia Nitrogen Analyzer is an online analytical device specifically designed for continuous monitoring of ammonium ion concentration in water bodies. It is widely used in wastewater treatment plants, waterworks, surface water monitoring, and industrial process control. Its core measurement principle is the Gas Sensitive Electrode (GSE) method, where a selective electrode reacts with ammonium ions in the sample, and the concentration value is ultimately output in the form of NH₄–N on the controller (sc1000).

Key Technical Features:

  • Wide Measurement Range: Covers three intervals: 0.05–20 mg/L, 1–100 mg/L, and 10–1000 mg/L, allowing flexible application in both low-concentration surface water and high-concentration wastewater scenarios.
  • Fast Response: 90% response time of less than 5 minutes, suitable for real-time monitoring of dynamic water quality.
  • High Precision and Reproducibility: Measurement error is less than ±3% or ±0.05 mg/L (for low ranges), ensuring reliable data.
  • Automation Capabilities: Features automatic calibration, automatic cleaning, and diagnostic functions, significantly reducing manual intervention.
  • Robust Design: Enclosure with an IP55 protection rating and made of UV-resistant ASA/PC material, suitable for harsh outdoor environments.
  • Modular Expandability: Enables data transmission and remote monitoring through the sc1000 controller, supporting single-channel or dual-channel modes.
    Thus, the Amtax SC combines high precision, low maintenance, and strong adaptability, making it a mainstream choice in the field of ammonia nitrogen online monitoring.

II. Installation and Calibration

1. Mechanical Installation

  • Mounting Options: Supports wall mounting, rail mounting, or vertical installation, with wall mounting being the most common. Choose a sturdy, load-bearing wall and ensure smooth routing of surrounding pipes and cables.
  • Weight and Load Requirements: The instrument weighs approximately 31 kg, and the bracket must support a load of ≥160 kg.
  • Installation Environment: Avoid strong vibrations, strong magnetic fields, and direct sunlight. Maintain an ambient temperature range of –20 to 45°C.

2. Electrical Installation

  • Must be performed by qualified personnel to ensure proper grounding and the installation of a residual current device (30 mA RCD).
  • Power is supplied by the sc1000 controller, with voltages of 115V or 230V. The use of 24V controller models is prohibited.
  • All piping and reagent installations must be completed before powering on.

3. Reagent and Electrode Installation

  • Reagent Preparation: Select standard solutions and reagents according to the measurement range. For example, use 1 mg/L and 10 mg/L standard solutions for low ranges, and 50 mg/L and 500 mg/L for high ranges.
  • Electrode Installation: Fill with electrolyte (approximately 11 mL), ensuring no air bubbles remain, and correctly insert the electrode into the electrolysis cell. Replace the membrane cap and electrolyte every 2–3 months.
  • Humidity Sensor: Must be correctly wired to prevent alarms triggered by condensation or liquid leakage.

4. Calibration Procedure

  • Calibration modes include automatic calibration and manually triggered calibration.
  • Set the calibration interval (typically once per day or shorter), and the system will automatically switch standard solutions for electrode correction.
  • After calibration, the system records key parameters such as slope, zero point, and standard solution potential to ensure long-term stable operation.

III. Startup and Operation

1. Startup Steps

  • Ensure all installations (piping, electrical, reagents, electrodes) are complete.
  • Connect the analyzer to the sc1000 controller and power on.
  • Initialize the system: Register the Amtax SC and sampling probe in the controller, execute the “Prepump All” function to fill the piping.
  • Allow a warm-up time of approximately 1 hour for the instrument, reagents, and electrodes to reach operating temperature.
  • Enter the sensor setup menu to confirm the measurement range, output units (mg/L or ppm), and measurement interval.

2. Normal Operation

  • LED Indicators: Green indicates normal operation, orange indicates a warning, and red indicates an error.
  • Measurement Interval: Adjustable from 5 to 120 minutes, depending on application requirements.
  • Data Viewing: The sc1000 controller displays real-time values, historical trends, and alarm status, and can upload data to a monitoring system via a bus interface.
  • Cleaning Function: Set up timed automatic cleaning to ensure the photometer, piping, and electrodes remain clean.

IV. Troubleshooting and Maintenance

1. Routine Maintenance

  • Appearance Inspection: Regularly check for damage to pipes and cables, and confirm the absence of leaks or corrosion.
  • Fan Filter: Clean or replace every 6–12 months to ensure proper heat dissipation.
  • Reagents and Electrodes: Replace reagents every 2–3 months, electrode membrane caps and electrolyte every 2–3 months, and electrodes every 1–2 years, as recommended in Table 5.
  • Cleaning Cycle: Depends on water hardness; typically perform automatic cleaning every 1–8 hours.

2. Common Faults and Solutions

  • Low/High Temperature: If the internal temperature falls below 4°C or rises above 57°C, the system enters service mode. Check the heating or cooling fan.
  • Humidity Alarm: Liquid detected in the collection tray; locate and repair the leak source.
  • Abnormal Electrode Slope: Check the membrane and electrolyte, replace the standard solution; if the issue persists, replace the electrode.
  • Weak Photometer Signal: Trigger cleaning; if unresolved, manually clean or contact a service technician.

3. Long-Term Shutdown and Storage

  • Flush the instrument with distilled water in a circulation mode to empty the pipes and reagent bottles.
  • Remove the electrode, clean it, and reinstall it in the electrolysis cell, keeping it moist during storage.
  • Install transport locks and store in a dry, frost-free environment.

4. Professional Repairs

  • Certain components (such as pumps, compressors, and main circuit boards) must be replaced by the manufacturer or authorized service personnel. Typical service lives: pumps 1–2 years, compressors 2 years, all covered under warranty.

V. Conclusion

The Hach Amtax SC Ammonia Nitrogen Analyzer is a stable and highly automated online monitoring device. It features a scientific principle, clear installation requirements, a straightforward operation process, and comprehensive maintenance methods. By strictly adhering to the user manual and this guide, users can ensure the long-term stable operation of the device, providing reliable data support for water quality monitoring and wastewater treatment process control. Correct installation, regular calibration, and maintenance are key to ensuring the instrument’s long-term stable operation. Users should strictly follow the safety specifications in the operation manual, regularly replace reagents and electrodes, and promptly address fault alarms to ensure measurement accuracy and extend the instrument’s service life.

Posted on

Troubleshooting Guide for Raycus RFL-P50QB Fiber Laser

1. Introduction

Raycus is one of the leading manufacturers of fiber lasers in China. Its RFL-P series pulsed fiber lasers are widely used in metal marking, welding, cutting, and surface cleaning.

From the nameplate you provided:

  • Model: RFL-P50QB
  • Output Power: 500W
  • Power Supply: 24VDC / Max. 14A
  • Structure: Main laser unit + fiber delivery cable + laser output head

In practice, common problems with this equipment are mainly related to power supply, fiber, cooling system, control signals, and the laser module.


2. Common Fault Symptoms

  1. No laser output at all
    • Fans running, but no laser beam emitted.
  2. Significant power drop
    • Originally 500W, now only 100–200W, insufficient for welding or cutting.
  3. Unstable output
    • Power fluctuates, beam spot unstable.
  4. Alarm indicators or error codes
    • Typical errors: over-temperature, fiber fault, module error.
  5. Output head contamination or damage
    • Lens blackened, spot distorted or doubled.

3. Troubleshooting Process

Step 1: Power Supply Check

  • Measure the input voltage:
    • Rated requirement: 24VDC, max 14A.
    • Use a multimeter; voltage must remain within 23.5–24.5V.
    • If voltage is too low, the laser cannot start or will output weak power.
  • Check power source:
    • Ensure power supply capacity is sufficient.
    • Tighten loose wiring to avoid overheating.

👉 Key point: Low voltage → no output; ripple noise → unstable laser.


Step 2: Control Signal Check

  • Enable signal:
    • The laser requires an enable signal from external control (CNC / PLC / marking card).
    • Verify connectors are not loose or oxidized.
  • PWM / analog signal:
    • Power control is typically via PWM or 0–10V input.
    • Use oscilloscope or multimeter to confirm correct waveforms.

👉 Key point: Missing signals → no laser; noisy signals → unstable output.


Step 3: Cooling System Check

  • Water chiller:
    • RFL-P50QB requires water cooling.
    • Confirm chiller is running, water temperature at 25 ±1 °C.
    • Ensure no bubbles in the pipeline.
  • Fans:
    • From your photo, the fan intake is dusty. Clean it.
    • Weak airflow → overheating alarm.

👉 Key point: Poor cooling → overheating shutdown.


Step 4: Fiber & Output Head Check

  • Fiber condition:
    • Look for bends, dents, or crushing.
    • Severe bending increases loss or causes permanent damage.
  • Output head (QBH collimator):
    • Inspect lens for black marks or burn spots.
    • Clean with isopropyl alcohol (IPA) and lint-free wipes.
  • Coupling condition:
    • Loose coupling → spot distortion.

👉 Key point: Dirty fiber head → reduced power; damaged fiber → no beam.


Step 5: Laser Module Check

  • Drive current:
    • If power is normal but no light, module failure is possible.
    • Requires factory repair.
  • Power measurement:
    • Use a power meter to test actual output.
    • If significantly lower than rated, the module is aging.

👉 Key point: Aged module → weak power; burnt module → no laser.


4. Common Faults & Solutions

SymptomLikely CauseSolution
No outputPower supply fault / no enable signalCheck 24V supply, verify control input
Power dropDirty fiber head / module agingClean fiber, replace module
Unstable beamPower ripple / cooling issueReplace power source, fix chiller
AlarmOverheat / fiber alarmCheck cooling system, fiber endface
Distorted spotBurnt output lensReplace or repair output head

5. Maintenance Guidelines

  1. Keep air vents clean – blow dust with compressed air.
  2. Replace cooling water regularly – use deionized water or dedicated coolant, change every 3 months.
  3. Clean fiber connectors – use 99% IPA alcohol and lint-free swabs.
  4. Avoid frequent plugging/unplugging of fiber heads.
  5. Stable power supply – use a UPS or voltage stabilizer.

6. Conclusion

The Raycus RFL-P50QB fiber laser is a robust industrial device, but it depends on stable power, proper cooling, clean fiber optics, and correct control signals to function.

From your photos and video, the most likely issues are:

  • Dust-clogged fan → overheating
  • Dirty or burnt fiber output head → power drop
  • Cooling water issues → overheat alarms

👉 Recommended sequence:

  1. Check power input.
  2. Verify cooling system.
  3. Clean fan and fiber head.
  4. Measure output with power meter.
  5. If still faulty → send to manufacturer.